Analytical Validation of a Highly Quantitative, Sensitive, Accurate, and Reproducible Assay (HERmark®) for the Measurement of HER2 Total Protein and HER2 Homodimers in FFPE Breast Cancer Tumor Specimens

Table 3

H2T and H2D assay validation summary. The parameters tested in the validation of the HERmark assay are summarized and the percentage passing is based on a 2-fold cutoff. NA indicates that a detailed pairwise comparison was not necessary and so therefore the 2-fold cutoff was not applicable. 95th percentile at 2.2 fold cutoff.

ParameterH2T Summary% Within Two-fold CutoffH2D Summary% Within Two-fold Cutoff

AccuracyPassNA (100)PassNA (100)
SensitivityPassNA (100)PassNA (100)
PrecisionPass100Pass99.5
ReproducibilityPass96.4Pass 91.81
Linearity: CEPass99.7Pass99%
Linearity: Section SizePass100Pass 92.21
Specificity: Triple Re-embedPass97.1Pass94.6
Specificity: IsotypePassNA (100)PassNA (99.1)