Within-run reproducibility (imprecision) performed on fresh blood samples.

MeanSDCV%Published SD and CV [7]

Sample 1
 IGC (×103/cu·mm)0.380.02656.96SD < 0.12, CV < 25%
 IG%3.570.329.01SD < 1.5, CV < 25%

Sample 2
 IGC (×103/cu·mm)7.870.00587.87SD < 0.12, CV < 25%
 IG%1.10.19.09SD < 1.5, CV < 25%