Clinical Study
Automated Measurement of Immature Granulocytes: Performance Characteristics and Utility in Routine Clinical Practice
Table 6
Within-run reproducibility (imprecision) performed on fresh blood samples.
| | Mean | SD | CV% | Published SD and CV [7] |
| Sample 1 | | | | | IGC (×103/cu·mm) | 0.38 | 0.0265 | 6.96 | SD < 0.12, CV < 25% | IG% | 3.57 | 0.32 | 9.01 | SD < 1.5, CV < 25% |
| Sample 2 | | | | | IGC (×103/cu·mm) | 7.87 | 0.0058 | 7.87 | SD < 0.12, CV < 25% | IG% | 1.1 | 0.1 | 9.09 | SD < 1.5, CV < 25% |
|
|