Clinical Study

Automated Measurement of Immature Granulocytes: Performance Characteristics and Utility in Routine Clinical Practice

Table 7

Estimate of agreement between different counting methods based on Passing-Bablock regression analysis for immature granulocytes.

Estimate95% CI

Automated IG% versus Manual IG%Slope0.940.83–1
Intercept−0.38−0.40–0.25

Automated IGC versus Manual IGCSlope0.930.86–1
Intercept−0.02−0.03–(−0.02)