Research Article

Validation of a Biofeedback System for Wheelchair Propulsion Training

Table 2

Validation of 𝐹 𝑥 𝑦 measurements.

Wheel angle (degrees)Actual force = 23.28 NActual force = 68.04 NActual force = 109.99 N
Measured force (N)Error (%)Measured force (N)Error (%)Measured force (N)Error (%)

023.11 ± 1.59−0.7466.55 ± 1.65−2.19108.68 ± 1.47−1.19
4523.75 ± 1.662.0267.05 ± 1.74−1.45109.17 ± 1.47−0.75
9022.94 ± 1.21−1.4866.26 ± 1.53−2.62108.25 ± 1.35−1.59
13524.04 ± 1.633.2467.39 ± 1.35−0.96109.45 ± 1.44−0.49
18023.85 ± 1.472.4367.08 ± 1.50−1.42109.23 ± 1.70−0.69
22523.38 ± 1.680.4166.65 ± 1.75−2.05108.92 ± 1.59−0.97
27022.75 ± 1.34−2.2865.84 ± 1.42−3.23107.78 ± 1.41−2.01
31522.40 ± 1.89−3.8065.51 ± 1.64−3.72107.43 ± 1.57−2.33

Measured values are mean ± 1 standard deviation.