Accuracy of Positive Airway Pressure Device—Measured Apneas and Hypopneas: Role in Treatment Followup

Figure 2

(a) Bland-Altman plot of PAP scoring versus manual scoring AI difference (PAP-scoring minus manual-scoring) by the mean. (b) Bland-Altman plot of PAP scoring versus manual-scoring HI difference (PAP scoring minus manual-scoring) by the mean. Note that all of the difference values were greater than 0, indicating that all PAP HI values were greater than the manual HI values. (c) Bland-Altman plot of PAP-scoring versus manual-scoring AHI difference (PAP scoring minus manual-scoring) by the mean. Note that all of the difference values were greater than 0, indicating that all PAP AHI values were greater than the manual AHI values.
314589.fig.002a
(a)
314589.fig.002b
(b)
314589.fig.002c
(c)