Clinical Study

SenseWear Armband and Stroke: Validity of Energy Expenditure and Step Count Measurement during Walking

Table 3

Agreement, standard error of measurement, and absolute percent difference by condition.

Variable, conditionICCSEMAbsolute percent difference

EE, SWA hemi versus SWA non-hemi0.6200.73NC
EE, SWA hemi versus metabolic cart0.5860.6817.9 (15.3)
EE, SWA non-hemi versus metabolic cart0.7020.4818.4 (13.3)
Steps, SWA hemi versus SWA non-hemi0.68297.92NC
Steps, SWA hemi versus SAM0.220151.539.3 (40.4)
Steps, SWA non-hemi versus SAM0.352132.3441.7 (38.0)

EE: energy expenditure; SWA: SenseWear Armband. SEM: standard error of measurement, units are kcal/min for EE, and steps for the steps variable. Values for absolute percent difference are mean (SD). NC: not calculated.