Determination of Iodate in Food, Environmental, and Biological Samples after Solid-Phase Extraction with Ni-Al-Zr Ternary Layered Double Hydroxide as a Nanosorbent

Table 2

Optimum conditions and analytical performance of the presented method for iodate determination.

Experimental conditionsUnit

Working pH5.5
Amount of Ni-Al-Zr LDH(mg)200
Sample volume(mL)50
Sample loading flow rate(mL min−1)3
Eluent concentration(mol L−1)2
Eluent volume(mL)1.5
Elution flow rate(mL min−1)1
Iodide concentration(mol L−1)0.1
Detection pH2.5
Final volume(mL)2.5
Wavelength(nm)352

Analytical parametersUnit

Linear range(μg mL−1)0.2–2.8
Intercept0.0389
Slope0.4184
Detection limita(μg mL−1)0.12
Correlation coefficient0.998
Relative standard deviation ( 𝑛 = 6 )b(%)2.5 (0.5)
Enrichment factorc20

aCalculated as three times the standard deviation of the blank signal.
bValue in parentheses is the iodate concentration (μg mL−1) for which the RSD was obtained.
cEnrichment factor calculated as the ratio between the volume of the initial aqueous solution and the final elution volume.