Research Article

Electrochemical Behavior of Quinoxalin-2-one Derivatives at Mercury Electrodes and Its Analytical Use

Table 5

Regression parameters of calibration curves y = ax + b (significance level α= 0.05, number of calibration points 𝑛 = 1 0 ).

AnalyteMethod (E P [V])c [μM]Electrolyte 𝑎 ± 𝑠 𝑎 [nA/μM] 𝑏 ± 𝑠 𝑏 [nA] [nA]rLOD [μM]

IDPV (−1.075)0.3–10.00.001 MNaOH2.74 ± 0.030.41 ± 0.210.99940.23
IDPV (−0.457)0.4–2.00.1 M HCl 1 2 . 3 9 ± 0 . 2 3 0 . 0 1 ± 0 . 3 0 0.99880.07
IIDPV (−1.010)0.3–10.00.001 M NaOH2.73 ± 0.030.25 ± 0.190.99950.21
IIIDPV (−1.100)0.1–1.0B.-R.-buffer pH 7.026.85 ± 0.21−0.17 ± 0.130.99980.01
IIIDPP (−1.080)0.1–1.0B.-R.-buffer pH 7.023.69 ± 0.13−0.18 ± 0.080.99990.01

EP versus SCE, r—correlation coefficient, LOD = 3 𝑠 𝑏 /a.