Research Article

Single-Channel Flow Injection Spectrophotometric Determination of Nickel Using Furildioxime in Micellar Solution

Table 5

Determination of nickel in Certified Reference Materials.

CRMComposition of CRMNickel found% recovery

Alloy50% nickel,
50% aluminium
49.9999.9

BAS (69b) stainless steel18.6% Cr,
9.35% nickel
9.3299.6

BCS 261, stainless steel0.083% C, 0.39% Si, 17.2% Cr, 13.08% Ni, 0.71% Ta & Nb,
0.66% Mn
19.0105