Research Article
Spectrofluorimetric Determination of Human Serum Albumin Using Terbium-Danofloxacin Probe
Table 3
Validation data of the proposed method for quantification of BSA, HSA, and HSA in human plasma.
| Parameters | BSA | HSA | HSA in plasma |
| Linear range (mol L−1) | 0.2 × 10−6–1.3 × 10−6 | 0.2 × 10−6–1.4 × 10−6 | 0.2 × 10−6–1 × 10−6 | Slope | 32.775 | 42.574 | 217.130 | Intercept | 3.9446 | 0.2742 | −14.6410 | Correlation coefficient | 0.9991 | 0.9990 | 0.9899 | Number of data points | 12 | 13 | 9 | LOD (mol L−1) | 8.7 × 10−8 | 6.2 × 10−8 | 8.1 × 10−8 |
|
|