Research Article

Spectrofluorimetric Determination of Human Serum Albumin Using Terbium-Danofloxacin Probe

Table 3

Validation data of the proposed method for quantification of BSA, HSA, and HSA in human plasma.

ParametersBSAHSAHSA in plasma

Linear range (mol L−1)0.2 × 10−6–1.3 × 10−60.2 × 10−6–1.4 × 10−60.2 × 10−6–1 × 10−6
Slope32.77542.574217.130
Intercept3.94460.2742−14.6410
Correlation coefficient0.99910.99900.9899
Number of data points12139
LOD (mol L−1)8.7 × 10−86.2 × 10−88.1 × 10−8