Research Article

Spectrofluorimetric Determination of Human Serum Albumin Using Terbium-Danofloxacin Probe

Table 5

Assay precision and accuracy of quality control samples.

Analyte nameConcentration (mol L−1)Intra-assay precision (RSD%)Inter-assay precision (RSD%) Accuracy (RE%)

BSA0.3 × 10−62.624.55−3.9
0.7 × 10−64.391.701.1
1.2 × 10−64.703.572.6

HSA0.3 × 10−63.194.074.4
0.7 × 10−62.794.623.1
1.2 × 10−62.002.60−0.4

HSA in plasma0.3 × 10−66.448.08−6.3
0.5 × 10−64.164.901.5
0.8 × 10−63.853.58−2.7