Research Article

Developing an Appropriate Digital Hearing Aid for Low-Resource Countries: A Case Study

Table 3

PDN-01B electroacoustics measurements.

IEC 60118-7Measured

HFA-OSPL90105 dB
Max OSPL90118 dB
HFA FOG 50 dB53 dB
Low frequency cutoff300 Hz
High frequency cutoff7.7 kHz
THD at 70 dB2.3%
Equivalent input noise38 dB
Current drain13 mA
Attack time1.3 ms
Release time382.9 ms
HFA-SPLITS105.0 dB
STS5.0 dB