Research Article

Design Optimization of Coronary Stent Based on Finite Element Models

Table 1

Optimization results.

WDS (mm)WTS (mm)WLS (mm) (mm) (MPa) Dogboning rateReduced by

Original stent0.280.280.2490.121.948 (LPV/LPC) 0.1452 (LPV/LPC)
1.9114 (SMPV) 0.0908 (SMPV)
0.0582 (LRD)
Optimal stent (LRD) 0.220.340.25680.13550.006189.52%
Optimal stent (LPV) 0.23670.220.20.11.7602 99.93%
Optimal stent (LPC) 0.24830.28810.20.11.9480.002798.14%
Optimal stent (SMPV) 0.32620.25820.20560.11.7491 99.89%