Abstract

Endo surgiclip instrument is the biomedical instrument that can be applied for endoscopic surgery to assist surgeons in homeostasis and secure mucosal gap surfaces during surgical operations. Since some clinic feedbacks show the surgiclip drop-off incidents which can potentially sever organ and tissue, the improvement of endo surgiclip instrument has been made in these years. Since few research papers were involved in the study of endo surgiclip instrument performance via mathematical modeling and computational simulation, currently some instrumental modifications are mainly based on clinic lab tests which prolong the improvement cycle and increase additional manufacturing cost. This paper introduces a new biomedical surgiclip instrument based on mathematical modeling, computer-aided simulation, and prototype testing. The analytic methodology proposed in this paper can help engineers in biomedical industry develop and improve biomedical instrument. Compared to the current conventional surgiclip instruments, this new surgiclip instrument can properly assist surgeon in surgical procedure with less operational force and no surgiclip drop-off incident. The prototype has also been built and tested. Both computational simulation and prototype testing show close results which validate the feasibility of this newly developed endo surgiclip instrument and the methodologies of mathematical modeling based computational simulation proposed in this paper.

1. Introduction

Biomedical/surgical instruments are renovated and technology-based devices requiring continuous improvement to guarantee safe use in surgery and be able to compete the challenging market [1]. The endo surgiclip instruments are commonly applied in surgical operations to repair biological defects, such as the anastomotic leakage in the gullet or gorge [2]. Recent researches indicate the flexibility and adaptability of endoscopic surgiclip instruments in curative and healing operation [3]. Endo surgiclip instruments can be applied in haemostasis using endoscopy to GI tract to repair the damaged tissues [4]. The biomedical/surgical instrument development is usually measured and controlled for its better functionality, proved feasibility, validated safety, improved reliability, and cost-effective productivity [5]. The functionality of biomedical surgiclip instruments is judged by the levels of bodily wound and tissue disturbance that biomedical devices might cause the patients [6]. The less-invasive biomedical/surgical instruments can be developed with advances of renovated surgical technique, improved biomedical technology, better engineering materials, and well-controlled manufacturing [7]. The design and development of biomedical/surgical instrument are to combine the engineering, biomedical and surgical technologies together to implement and improve medical diagnosis, clinic treatment, and surgical operation [8]. In developing the biomedical/surgical instruments, the interdisciplinary scientific knowledge is required including mathematics, engineering, chemistry, electronics, mechanism, and physics [9]. Some recent surgical and clinic feedbacks show the surgiclip drop-off incidents due to malfunction in endo surgiclip instrument causing severe bodily disturbance, tissue trauma, and organ damage during surgery [10]. To eliminate the above defects, the current endo surgiclip instruments must be modified and improved. In this paper, the endo surgiclip instrument modification is implemented by establishing the mathematical modeling of systematic mechanism and performing computational simulation based on predefined mathematical model. To verify the analytic solution, the prototype of this new biomedical/surgical instrument has been built and tested. Both mathematical modeling/computational simulation analysis and prototype experiment show closed results which validates the feasibility of research methodology proposed in this paper.

2. Newly Developed Biomedical/Surgiclip Instrument

This newly developed biomedical/surgiclip instrument, with many jointed moving parts that provide the desired instrumental motion, is shown in Figures 1 and 2.

The instrumental linkage mechanism controls the open and close function of tip jaw and spring unit keeps trigger system returning to the original position after jaw tip released. To function properly in surgical operation, 20 lbf forces should be applied to the jaw tip through the linkage mechanism in order to completely form the surgiclip to secure the body tissues. The materials for jaw tip, pusher bar, and slider are ASTM A304L containing 1.38% carbon content with coated Chromium to prevent these parts from carbon oxidation. One pusher bar end is jointed to the linkage mechanism and another end is connected to the jaw tip. When pusher bar at jaw tip end moves forward and passes the jaw external slop, the perpendicular force is transformed onto the jaw tip in order to form the surgiclip. To keep jaw tips normally open, an expansive spring is installed between the jaw tips to keep them open when trigger handle is released.

3. Mathematical Modeling and Computational Simulation

Figures 3, 4, and 5 show jaw external geometry, trigger and slider linkage system, and pusher and jaw tip mechanism respectively. The force required to overcome frictional forces at points and along jaw external slop is

Therefore the total force required is

Here, is the force required to close jaw tips. When jaw tips are fully open, the jaw external slop angle is 16°, shown in Figure 3.

Figure 4 displays the trigger and slider linkage system and Figure 5 shows the pusher bar and jaw tip mechanism.

The force is required to drive pusher bar as slider moves forward. Stiffness of compression spring and torsion spring are 0.465 and 0.66, respectively.

Force required to overcome friction at point is

Force required to overcome friction at point is

Total force required to overcome friction at points and is

Force required to overcome friction at point is

Force required to overcome friction at point is

Total force required to overcome friction at points and is

So the total force required to overcome frictions from points , , , , , and is

The force angles at contact points and change along jaw external slop and force angles at contact points , , , and varies with different trigger closure angle. The force required to handle the trigger is

For compression spring shown in Figure 6, the compression spring rate is and the maximum shear force in compression spring is

The torsion spring, installed at the back of trigger handle, can be applied to move trigger back to its original position after trigger is released. For torsion spring displayed in Figure 7, the force loaded on torsion spring is

The above complex calculation of forces in (11) at different force angles can be performed through computational simulation based on predefined mathematical models.

The trigger design needs considering handle body strength to keep better instrumental function, ergonomic contour features to prevent surgeon hands from fatigue, and less operational force to ease surgeon in surgical operation.

Figure 8 displays the force diagram and Figure 9 shows the geometric diagrams of trigger in its start (jaw tips fully open) and end (jaw tips fully closed) positions. Based on Figure 9, the distance of L-J in vertical direction is

As the surgeon applies force to the instrument trigger, the other end of trigger handle will drive the instrumental pusher bar forward to load surgiclip into jaw tip track. The pusher bar stays inside jaw tip but a small front portion of surgiclip moves outside jaw tip before reaching the tissue target. The surgiclip will further move and secure body tissue as the surgeon continuously operates the instrument.

Based on the above mathematical modeling (11)–(15), the instrumental operation force can be determined through computational simulation and the analytic result is shown in Figure 10.

The computational simulation indicates that the maximum operation force of this new biomedical instrument to fully form the surgiclip is 3.725 lbf which is lower than 4 lbf operation force in current traditional surgiclip instruments due to improved driving mechanism in this new instrument design. This improvement will benefit the surgeons in their surgical operations and clinic treatment applications.

The computer-aided simulation is applied to perform optimal design on this endosurgical instrument. The combinations of different linkage angles and length ratio are introduced in the computational simulation to determine the above combination of linkage angle and length ratio that is best for this biomedical instrument design.

4. Prototype Testing of this New Biomedical Instrument

The prototype of this new biomedical instrument has been built and tested to verify the mathematical modeling based computational simulation. The experimental result is shown in Table 1.

The experimental results of this new biomedical instrument show the average maximum operational force to fully form the surgiclip is 3.7267 lbf which is very close to the maximum operational force of 3.725 lbf by mathematical modeling based computational simulation. This validates the credibility of the mathematical modeling proposed in this paper. Also, the prototype testing indicates that no surgiclip drop-off incident during testing due to improved instrument mechanism design.

5. Future Improvement on this New Biomedical Instrument

The future improvement will include further simplifying the instrument design to lower manufacturing cost, investigate possible design solution to further decrease instrument operation force, reduce linkage numbers to bring down friction loss, and modify body/trigger ergonomic contour to further trim down the risk of hand fatigue. The prototype of further modified instrument will be built and tested in the lab and will also be sent to the clinic field for further evaluation thereafter.

6. Conclusion

The improvement of biomedical instruments can efficiently help doctors and surgeons in the medical treatment and surgical procedure. This new biomedical/surgical instrument is designed to prevent endo surgical instrument from surgiclip drop-off incident to avoid severe body tissue injury in surgical procedure and prototype experiment indicates no surgiclip drop-off incident in medical treatment. The mathematical modeling based computational simulation shows lower operational force required in this new instrument design if compared to the current existing surgiclip instruments and this improvement can ease surgeon in their future surgical operation. Both mathematical modeling based computational simulation and prototype experiment show very close results which validates the credibility of analytic methodology proposed in this paper.