Abstract

State-of-the-art transtibial prostheses provide only ankle joint actuation and thus do not provide the biarticular function of the amputated gastrocnemius muscle. We develop a prosthesis that actuates both knee and ankle joints and then evaluate the incremental effects of this prosthesis as compared to ankle actuation alone. The prosthesis employs a quasi-passive clutched-spring knee orthosis, approximating the largely isometric behavior of the biological gastrocnemius, and utilizes a commercial powered ankle-foot prosthesis for ankle joint functionality. Two participants with unilateral transtibial amputation walk with this prosthesis on an instrumented treadmill, while motion, force, and metabolic data are collected. Data are analyzed to determine differences between the biarticular condition with the activation of the knee orthosis and the monoarticular condition with the orthosis behaving as a free-joint. As hypothesized, the biarticular system is shown to reduce both affected-side knee and hip moment impulse and positive mechanical work in both participants during the late stance knee flexion phase of walking, compared to the monoarticular condition. The metabolic cost of walking is also reduced for both participants. These very preliminary results suggest that biarticular functionality may provide benefits beyond even those of the most advanced monoarticular prostheses.

1. Introduction

The loss of a leg below the knee can have a formidable impact on one’s quality of life [13]. The widespread, passive ankle-foot prostheses on the market today provide only a rudimentary approximation to the function of a human ankle joint [4, 5]. Instead of providing net mechanical work to the wearer during walking, these passive devices act at best in a spring-like manner; they can only provide as much mechanical energy return as is provided to them by the wearer, and they do not provide the articulation normally seen in the biological ankle-foot complex during walking. As evidence of this technological limitation, transtibial amputees display a variety of pathological features of their walking gaits. Specifically, transtibial amputees naturally select a 30–40% slower walking speed than those without amputation, and when walking at the same pace as a nonamputee, these amputees require 20–30% more metabolic power than their nonamputee counterparts [4, 68]. Further, these amputees exhibit increased levels of hip positive power during late stance phase. This increased hip power may be a compensatory response to lack of calf muscle function and could contribute to the aforementioned increase in metabolism while walking [4].

In the last several years, robotic advances in prosthetic technology have led to the introduction of powered ankle-foot prostheses (EmPower, BionX Medical Technologies, Inc., Bedford, MA), which, unlike the passive conventional devices, provide levels of mechanical work comparable to those provided by the human ankle-foot complex. As a result of this functional improvement, many of the aforementioned gait pathologies have been drastically reduced; amputees using the powered prostheses have preferred walking speed, metabolic cost at a given speed, and contralateral limb impacts that are not significantly different from those of nonamputees [9, 10]. These improvements are thought to stem from the propulsive net mechanical work provided by these devices to the wearer [9], as this propulsion helps to redirect the center-of-mass, thereby reducing collisions of the contralateral limb [1114]. This ankle propulsion may also assist swing-initiation of the affected-side limb [1518].

These new prosthetic devices are, however, limited to emulating the function of the ankle-foot complex alone and consequently cannot restore the full function of the powerful gastrocnemius muscle. The gastrocnemius provides not only a plantar flexion moment at the ankle, but also a flexion moment at the knee. Without the knee-flexing function, compensatory mechanisms are required. Indeed, transtibial amputees exhibit higher hamstring muscle activity during level-ground walking than nonamputees [19], possibly as an attempt to stabilize or flex the knee in place of the nonfunctional gastrocnemius muscle. This higher muscle activity is still apparent when amputees walk with the powered ankle-foot prostheses, indicating that a monoarticular intervention at only the ankle-foot complex may not be sufficient to restore biological function. It is possible that this pathological muscle activity has detrimental effects on amputee gait. However, little work has been done to develop devices for restoring this missing gastrocnemius functionality.

Since those with transtibial amputation maintain their biological knee joints, a direct assistance of these joints requires the use of an exoskeletal device such as an orthosis that provides assistance in parallel to the existing joint. A variety of robotic orthoses have been developed to assist human knee joints for locomotion. Many of these devices utilize active components such as electric motors [2022] or pneumatic actuators [2326], but these elements tend to require a large power source or otherwise require off-board tethered actuation, limiting the scope of their applicability. Alternatively, some researchers have developed quasi-passive knee exoskeletons that do not provide net mechanical power to the wearer but instead require at most the minimal input power to operate the electronics and mechanical components [2730]. In one such study, a spring-loaded, custom toothed clutch was designed for assisting extension moment during human running [29]. This device locked to support the knees of nonamputee participants during the stance phase and unlocked to be a free joint during the swing phase. Using this quasi-passive device, large knee moments of 190 Nm could be supported with an exoskeletal joint weighing only 710 grams. This, and other exoskeletons have been built to assist with knee extension in nonamputees, but no such devices to date have been built for knee flexion assistance, particularly for the gastrocnemius, in those with transtibial amputation.

In order to develop an assistive device for the gastrocnemius, it is important to first understand how this muscle functions. In vivo ultrasonography shows that the gastrocnemius muscle fascicle length is largely isometric during the early and mid-stance phases of level-ground walking [31], indicating that the passive tissues such as tendons are responsible for much of the power delivery from the gastrocnemius. Hence, the gastrocnemius may be approximated by a clutched spring, with the spring representing the compliant tendinous structures, and the clutch representing the isometrically acting muscle fibers. Indeed, a model of human gait by Endo and Herr with only spring-clutch structures at the knee joint was able to achieve human-like metabolic economy, while capturing the dominant kinetic features of human walking [3234]. It is therefore possible that a physical clutch in series with a spring can provide much of the missing knee function of the gastrocnemius in transtibial amputees.

Researchers have begun to study the effects of a robotic device at the affected-side knee to provide the knee flexion moment of the missing gastrocnemius. In the first such study [35], a quasi-passive knee orthosis, dubbed an artificial gastrocnemius (AG) was built, based on the spring-clutch representation of the gastrocnemius in the aforementioned Endo-Herr model. The AG was physically realized as clutchable rotary spring at the orthosis knee joint. Two copies of this AG, worn along with a powered-ankle-foot prosthesis (PAFP), was tested on both legs of one bilateral transtibial amputee, with promising results: the amputee’s metabolic energy expenditure was reduced with the AG-PAFP condition, compared to the amputee walking at the same speed with only conventional leaf-spring-type ankle-foot prostheses. However, since the PAFP was not tested independently of the AG, it is not clear what incremental effects the AG had over the PAFP alone. A need, therefore, exists to evaluate the incremental biomechanical and metabolic effects of the AG.

In this study, we design and evaluate a new version of the AG unit, based on the exoskeletal knee clutch design [29]. We evaluate the effects of this new AG-PAFP on the walking gait of transtibial amputees. We focus on the late stance knee flexion phase of walking, during which the knee flexes to prepare for the swing phase, so to detect improvements on the pathologies thought to be most likely affected by the lack of calf muscle function [4]. We expect that the energy stored in the spring of the AG early in the gait cycle will be returned during late stance knee flexion to replace positive muscle work. Specifically, we hypothesize that including a clutched-spring AG at the affected-side knee joint of transtibial amputees in conjunction with a PAFP will reduce both biological flexion moment impulse and positive mechanical work of the affected-side knee and hip joints during late stance knee flexion, as compared to the same conditions with only the PAFP. We further hypothesize that these changes will provide a corresponding reduction in metabolic cost of walking. We expect that this metabolic reduction will stem from reduced joints moments and net work in the affected knee and hip joints. We evaluate these hypotheses by analyzing the kinematics, kinetics, and metabolism of two transtibial amputees walking with the AG-PAFP combination on an instrumented treadmill. We compare these data between the two conditions where the AG is active versus acting as a free-joint.

2. Methods

Two devices were used for the intervention in this study. A powered ankle-foot prosthesis provided ankle function in the sagittal plane, and a quasi-passive artificial gastrocnemius (AG), that was a clutched-spring joint, was mounted on a knee orthosis on the affected-side knee. Together, these two devices represented a biarticular transtibial prosthesis that could actuate both joints independently.

2.1. Powered Ankle-Foot Prosthesis

An EmPower powered ankle-foot prosthesis (BionX, Bedford, MA) was used as the prosthesis for all clinical trials. This prosthesis had the capability of providing positive net power at levels comparable to the human ankle-foot complex [9]. At the core of the prosthesis was a series elastic actuator, comprising a brushless DC motor, ball screw, and carbon fiber leaf spring. For dorsiflexion angles, a rigid hard-stop engaged, both to reduce the torque required by the motor in dorsiflexion and to act as a safety feature in the event of a power failure. The total mass of the prosthesis and battery was 1.8 kg.

The prosthesis controller employed a positive force feedback strategy that served to approximate the biological muscle dynamics and neural reflexes. A wireless communication link enabled real-time tuning of the control parameters.

2.2. Clutch-Spring Joint

The AG, shown in Figure 2, was a knee orthosis, comprising a pair of polycentric hinges that connected a thigh cuff to an 1/8′′-thick aluminum bracket made of aluminum sheet stock (6061-T6). The bracket connected to the base of a participant’s socket, between the prosthetic pylon attachment and the socket. The proximal side of the brace was strapped to the participant’s thigh as a typical knee orthosis. A custom, dog-tooth rotary clutch spring with series compliance was attached lateral to the knee, in parallel with the hinges.

This clutch-spring joint was based on a previous design [29] but had the addition of series compliance and reversed the direction of the clutch teeth so that the teeth engaged to provide a knee flexion moment, rather than an extension moment to the wearer. In addition, our device lacked a planetary gear train that was present in the previous work. The machined clutch components were made of 6061-T6 aluminum alloy and included two rings of dog teeth, brought together by the action of a solenoid (LT8x9, Guardian Electric, Woodstock IL). As in Figure 1, the clutch controlled the relative rotation of the Housing to the Rotating Clutch Ring. A solenoid was built into the middle of the joint and translated in the medial-lateral direction under action of its actuation power and a return spring. When the solenoid was energized, the Translating Clutch Ring was engaged with the Rotating Clutch Ring. The Translating Clutch Ring constrained the rotation of the Rotating Plate that, in turn, was connected to the Housing through two tangential linear compression springs. Thus, when the solenoid was energized, the Housing was coupled to the Rotating Clutch Ring through these springs. When the solenoid was inactive, the return spring separated the Translating Clutch Ring from the Rotating Clutch Ring and, hence, they were then free to rotate with respect to each other. The return spring stiffness was chosen experimentally so that when the clutch developed more than 0.5 Nm of torque, the tooth friction prevented the two rings from separating. This tooth-binding effect acted as a safety mechanism and was included as a part of the control algorithm, as described in Section 2.8. This aforementioned friction maintained the clutch in a clutched state until the torque level dropped back to the safe value of 0.5 Nm. As a result, the bulk of the stored spring energy could only be released gradually into the wearer’s leg, rather than suddenly into the mechanism.

The Housing of the clutch was bolted to the thigh cuff of the knee orthosis. The Rotating Clutch Ring connected to a distal output link with a linear ball bearing and radial ball bearing, acting in series. These bearings accommodated for the kinematic difference between the polycentric hinges of the orthosis and the single-axis rotation of the clutch joint. Thus, only forces tending to flex or extend the joint could be transmitted through this set of bearings.

2.3. Sensing

The AG had onboard sensing for use in the control algorithm. Knee angle of the AG was measured using a 10 kOhm rotary potentiometer. Knee moment provided by the spring-clutch element to the wearer was estimated by measuring the deflection of one of the two identical tangential springs with an 8 kOhm linear potentiometer. The voltage output from the potentiometer was scaled with an experimentally determined gain to provide an estimate of torque applied by the AG to the wearer. This gain was found by applying various tangential forces to the distal end of the brace using a force gauge (gauge moment arm = 28 cm), while the clutch was engaged. With this force and the known moment arm of the gauge, the applied knee moment was computed and compared to voltage readings from the linear potentiometer to achieve a scaling factor for conversion from voltage to Nm/rad. The resulting measured spring-clutch knee torque reading is subsequently referred to as the AG knee moment. A similar calibration procedure was performed using the AG angle-sensing rotary potentiometer and an infrared camera system (model T40s, Vicon Motion Systems Ltc, Oxford, UK). In this calibration procedure, reflective markers were placed on the center of the joint and on the distal and proximal ends of the orthosis. Marker-based joint angle was computed by obtaining the relative angles of lines connecting the distal and proximal markers to the joint center marker. These angles were then compared to voltage readings from the angle-sensing potentiometer. Prosthesis-side ground contact was detected using a resistive pressure sole footswitch (model: FSW, B&L Engineering, Santa Ana, CA), inserted into the shoe between the prosthetic foot and the insole.

2.4. Failure Analysis

Unlike the titanium toothed rings in the exoskeletal clutch [29], the clutch in the present study was made from aluminum. Thus, it was necessary to reevaluate the load-bearing limits of the teeth. Finite element analysis (FEA) was performed using Solidworks (DS Solidworks Corp, Waltham, MA) to ensure this clutch could support the required knee moments. Tangential loads were simulated symmetrically on the tips of all teeth to simulate worst-case loading scenario.

2.5. Modeling and Spring Stiffness Selection

The selection of the spring constants for the tangential springs allowed for the control of rotary stiffness of the clutch-spring element. It was desirable for the stiffness of the clutch-spring joint to be such that the spring-clutch behavior of the artificial gastrocnemius would most closely reproduce the gastrocnemius muscle behavior of a healthy gastrocnemius. To this end, walking data from healthy nonamputees was fed as input to a modified version of the Endo-Herr sagittal-plane human leg model [3234] during the development of the artificial gastrocnemius in [35].

2.6. Control

The quasi-passive artificial gastrocnemius produced no positive net mechanical work to the wearer but still required a control system to determine the appropriate times to engage and disengage the clutch. This controller took knee joint angle and stance information as input and engaged the clutch appropriately through each gait cycle. The clutch was controlled to engage at maximum stance knee flexion angle, enabling the spring to store energy during the subsequent knee extension and flexion into swing phase, as shown in Figure 3.

2.7. Control Electronics

The computer platform for controlling the AG was a commercial single-board computer (model: Raspberry Pi Version B, Raspberry Pi Foundation, Cambridgeshire, UK). The computer was equipped with an 800 MHz ARM11 processor, 512 MB SDRAM, with Linux Debian. The system was powered by a 6-cell lithium polymer battery (nominal voltage of 22.2 V).

2.8. Control Algorithm

High-level control was implemented using a finite state machine, implemented in Python. The gait cycle was divided into four states, shown in Figure 4: Swing; Early Stance; Clutched; and Late Stance.

The Swing state was triggered from any other state when the affected-side foot left the ground, as detected by a drop in footswitch signal, FSW, to less than a fraction, of the maximum possible signal. During Swing, the clutch was disabled, allowing the knee to swing freely.

The Early Stance state was triggered from the Swing state at foot contact of the affected-side with the ground, defined as FSW increasing to the stance threshold, , provided that the time elapsed in the current state, , was at least the minimum time required for the Swing state, . During the Early Stance state, the clutch was disabled, and the controller monitored the knee angle of the brace for maximum stance knee flexion, at which point the Clutched state would be engaged. A least squares algorithm, similar to the one used previously [29], continually predicted the time remaining before maximum knee flexion angle. This prediction provided time to initiate the engagement of the clutch, so that the clutch would be fully engaged as close as possible to the time of maximum knee flexion. In addition to this knee flexion detection algorithm, two safety features were in place to ensure that the motions detected were resulting from a walking gait. The Clutched state could only be enabled if the following two conditions were also met: the knee was flexed a minimum angle of from the angle at heel strike, ; the maximum knee flexion angular velocity, , measured during the Early Stance state was at least . The values and were experimentally determined during early testing as the lowest values that successfully prevented false-triggering of the Clutched state outside of a steady, level-ground walking gait (Table 1).

The Clutched state served to activate the clutch near the maximum knee flexion during stance phase of walking. This state was activated when the following criteria were met: the maximum stance flexion angle was predicted to occur within the time of the current time, the knee angle was flexed more than a threshold, , and the maximum knee flexion angular velocity, , during the Early Stance state exceeded a threshold . Conditions and were used to differentiate a walking gait with slow, nongait motions, the latter of which did not warrant activation of the clutch.

The Late Stance state served to turn off the clutch after it engaged. Once the clutch spring began to develop force, the clutch teeth would bind, preventing the clutch from disengaging until the spring force dropped sufficiently. Therefore, the clutch solenoid was deactivated by entering the Late Stance state when the time elapsed in the Clutched state, , exceeded the clutch timeout threshold, .

In order to maximize spring energy storage and return, it was desirable to engage the clutch as near as possible to the moment of peak knee flexion in the Early Stance state. In fact, the clutch solenoid needed to be engaged slightly before the desired clutching time, as to account for the time required to close the gap between the two sets of clutch teeth. To achieve the necessary prediction of peak stance flexion, a look-ahead algorithm was used.

First, the knee angle during the Late Stance state was approximated as parabolic, given this assumption, the location of the vertex of the parabola may be found by performing a linear fit to the knee angular velocity data, via running sums, and solving for the zero-crossing. The algorithm used here was similar to one described by [29]. However, the earlier algorithm assumed a fixed time step that was not applicable with the system in this study, as the computer platform was not a truly real-time system. Hence, the algorithm proposed here did not presume a fixed time step.

The parameter that minimizes the error in the least squares sense of linear model to time-series data iswhere the matrix has elementsand is the th element in the discretely sampled vector. Expanding (1) yieldswhere all summations in (3) span from 1 to . For the task of estimating the angular velocity over time, the values of were timestamps, and the values of were knee angular velocity values. For the window of size , the most recent values of and were maintained in a queue. At each new timestep, the oldest values of and were popped from the queue, while the current and values were added to the queue, maintaining the 0th element as the first in the queue. This method allowed the sums to be updated each timestep without needing to store all values of the computed sums. Finally, the estimated time of the angular velocity zero-crossing was . It was found experimentally that the minimum error in clutching time between the initiation of the development of clutch torque and the peak knee flexion angle occurred when the clutch was engaged 50 ms prior to the predicted maximum knee angle. As a result, the time delay parameter, , was set to 50 ms.

2.9. Experimental Protocol

Two participants with below-knee amputation were involved in this study (Table 2). Both participants had right-side unilateral amputation and were of generally good health. The clinical evaluation was conducted at MIT (Cambridge, MA) and was approved by MIT’s Committee on the Use of Humans as Experimental Subjects (COUHES). Each participant provided written, informed consent that was obtained before data collection was initiated.

An infrared camera system (model T40s, Vicon Motion Systems Ltc, Oxford, UK) was used to track the three-dimensional motion, recorded at 100 Hz, of reflective markers, placed at 47 anatomical locations on the participants’ bodies, based on the Helen Hayes marker model. Ground reactions forces and center of pressure locations were measured using a dual-belt instrumented treadmill (Bertec Corporation, Columbus, OH) with a sampling rate of 1 kHz. The net metabolic cost of walking during each condition was estimated using standard open-circuit gas exchange techniques (model: K4b2, COSMED, Rome, Italy).

At the beginning of each session, the participant was asked to don the powered prosthesis in place of their conventional prosthesis. The knee orthosis of the AG was then affixed to the prosthesis and donned by the participant.

For each participant, the prosthesis controller’s power setting was adjusted using the commercial tuning app as the participant walked over a treadmill at 1.25 m/s so as to achieve net work from the prosthesis per step that was within one standard deviation of the mean for nonamputees walking at the same walking speed [36]. It was verified that the level of prosthesis net work remained within the desired range for both walking conditions for a given participant (0.045 to 0.16 J/kg) and that this level of net work stayed reasonably consistent across walking conditions for each participant.

A short time period of approximately 15 minutes was given to ensure that both the prosthesis and the artificial gastrocnemius orthosis were functioning properly. This acclimatization was evaluated by verifying that the clutch was engaging and disengaging at appropriate points in the gait cycle (as in Figure 3) and that the PAFP net mechanical work was reasonable as compared to biological values.

Both the powered prosthesis and artificial gastrocnemius were worn for all trials. The participants were asked to perform one standing trial to measure standing metabolism. Walking trials were performed on the treadmill at a speed of 1.25 m/s.

Two walking conditions were tested: a baseline condition (Baseline) in which the AG acted as a free-joint at the knee with the spring-clutch disabled, and an active condition (Active), in which the AG was controlled as a clutched spring at the knee with the described control algorithm. For both conditions, the powered ankle-foot prosthesis was active. The Baseline condition represented a monoarticular transtibial prosthesis, as the knee joint was a free-joint when the clutch was inactive. Yet the mass distribution of the device was identical to the Active condition. Thus, a direct comparison could be made to determine the incremental effects of the clutched-spring knee joint.

2.10. Data Processing

Fourth-order Butterworth filters were used to filter the marker position and ground reaction force data with 6 Hz and 25 Hz cutoff frequencies, respectively. The marker and force data were postprocessed through the SIMM (Musculographics Inc., Evanston, IL) inverse dynamics module to produce total joint moments and angles in three dimensions. However, only sagittal-plane dynamics were considered. Affected-side biological knee moment contribution was computed by subtracting the AG knee moment from the total knee moment estimated from the SIMM-based inverse dynamics.

Gait events were determined using vertical ground reaction force data from the embedded force plates. Approximate event timing was found by determining the times when the force increased beyond a 40 N threshold. Exact heel strike and toe-off times were found by progressing backward and forward in time, respectively, until the force value dropped to zero. Data were then cut to gait cycles based on heel strike times and resampled to 101 points. Gait cycles were discarded for the beginning and end of each trial, during the speed transients of the treadmill. Gait cycles in which the stride times were below 0.7 seconds or above 1.3 seconds or in which a foot crossed the midline of the treadmill were also discarded.

Joint powers were computed as the product of joint moments and joint velocities from SIMM-derived joint kinematics, where positive power was defined as that produced by the joint on the environment. Positive joint work during late stance knee flexion was computed as the positive contribution of the time-integral of joint power from maximum stance knee extension to toe-off. This region of the gait cycle was chosen for analysis because, as the knee flexes from the maximum extension angle, it provides an opportunity for the AG to provide positive power to the wearer. Joint flexion moment impulse was computed using the same integral for joint flexion moment. Net prosthesis work was computed by integrating the SIMM-derived joint torque over joint angle for the entire gait cycle.

Metabolic cost for each walking speed was computed by taking average oxygen and carbon dioxide data over a two-minute window at the end of each six-minute trial. The metabolic power was computed using the equationwhere is the metabolic power in Watts, is the volume flow rate of oxygen inhaled, is the volume flow rate of carbon dioxide exhaled, and and are constants with values from literature [37], given as = 16,580 W/L and = 4,510 W/L. The above equation is only valid for conditions when the metabolism is primarily aerobic. As a verification of this condition, the respiratory exchange ratio (RER), defined as , was monitored, and only metabolic results with RER values less than 1.1 were considered.

3. Results

3.1. Mechanical Design

The mechanical parameters of the design are summarized in Table 3. As in previous work [29], the teeth were manufactured using a contour milling process, and thus, the tooth spacing was limited by the 0.4 mm radius of the mills. Therefore, the tooth spacing was constrained to 4 degrees, or 90 teeth per revolution. The total mass of the orthosis, including battery and electronics, was 1.9 kg.

The aluminum teeth used in this design do not yield for applied knee flexion moments near to the expected values of 25 Nm [38] but start to yield at 35 times this expected load (Table 3). The tooth stress at this failure load condition is shown in Figure 5.

3.2. Power Consumption

Given the solenoid specifications (22.2 V, 109 Ohms resistance), the power draw was 4.9 W. Assuming the maximum clutch-on time for each gait cycle of 0.4 ms and a typical cadence of 0.9 strides per second, the power consumption of the solenoid was 2.2 W. With the 1.1 W power draw of the Raspberry Pi, the average total power consumption of the device was 3.3 W during a typical walking gait.

3.3. Joint Spring Selection

The spring stiffness of the clutched-spring gastrocnemius element, derived from the modeling, was 92.9 Nm/rad [35]. To approximate this rotary stiffness about the knee joint, the two tangential linear compression springs were selected with spring constants of 51 N/mm. These springs acted on the joint with a moment arm of 32.1 mm, causing the equivalent joint rotary stiffness to be 105 Nm/rad.

3.4. Clinical Pilot Results
3.4.1. Acclimatization

In the time allotted for acclimatization, participants could walk with the AG and PAFP. Participants could walk comfortably with the AG, and the ankle prosthesis net work was appropriate for normal walking (see Table 4).

3.4.2. Ankle Prosthesis Net Work

As shown in Table 4, the net work produced by the powered ankle-foot prosthesis was within the desired range from literature [36].

3.4.3. Affected Knee Kinetics

Affected-side knee kinetics for both participants are shown in Figure 6 and summarized in Tables 5 and 6. Of particular note, the biological component of knee flexion moment impulse and the biological knee positive mechanical work during late stance knee flexion phase were both reduced by large percentage changes when subjects walked with the Active condition of the AG, compared to the Baseline condition.

3.4.4. Affected-Side Hip Kinetics

Hip flexion moment is shown in Figure 7, and hip flexion moment impulse values are summarized in Table 7. Hip net mechanical work values for both participants are shown in Table 8. As with the knee joint, both hip flexion moment impulse and positive mechanical work during late stance flexion were decreased slightly for both participants.

3.4.5. Metabolism

The net metabolic power (with the power required for standing subtracted out) is tabulated in Table 9. Both participants displayed small percentage reductions in metabolic cost when walking with the Active condition, as compared to the Baseline condition.

4. Discussion

Our novel design of a quasi-passive artificial gastrocnemius has been tested in our pilot study and found to be mechanically functional for the task of providing a flexion assist moment to the affected-side biological knee joint. The device exceeds the torque requirements while requiring only a small amount of electrical power. These properties make this a device a prime candidate for experiments for those with transtibial amputation, with the end goal of improving walking gaits for this population.

The preliminary findings from this study indicate that some measures of gait may be especially affected by the intervention. The pilot data support the hypothesis that this artificial gastrocnemius can reduce biological knee flexion moment and biological positive work of the affected-side knee, compared to a similarly weighted monoarticular prosthesis, during late stance knee flexion. Despite the short amount of time to get acclimated to the knee orthosis, the participants reduced their biological knee moment profiles, thereby allowing the AG to take over some of the kinetic loads. This behavior is consistent with other studies involving exoskeletal interventions at ankle [39] and hip joints [40]. The reduction of biological knee moment is beneficial from the standpoint of wearable device design, since it means that wearers of the AG were quickly able to replace biological function with that from the device.

Also as hypothesized, the affected-side hip flexion moment impulse and hip positive work of both participants were reduced during late stance knee flexion. These reductions are likely a result of the energy return of the spring, which served to help flex the hip. Although small, this effect may possibly compensate for the otherwise increased hip power exhibited by transtibial amputees [4].

Given the kinetic results, it is not surprising that slight metabolic improvements were demonstrated as well. Reductions in net positive biological mechanical work should correspond to reductions in concentric muscle work, since passive tissues, by definition, cannot generate net positive work. This reduction in muscle work, in turn, would be expected to reduce metabolic cost [41, 42].

5. Conclusions and Future Work

Despite the advances of today’s monoarticular transtibial prostheses, their limitations still manifest in pathological gaits. This study builds on the hypothesis that dominant pathologies are caused by a lack of biarticular gastrocnemius function. Our approach of restoring this biarticular component via an artificial gastrocnemius has shown preliminary but promising results toward the improvement of transtibial prosthesis efficacy.

The quasi-passive nature of this novel artificial gastrocnemius means that the design could exclude heavy components like large batteries and motors. Yet still, this device has the capability to produce biological-levels of knee moments for assisting those with transtibial amputation. The low power draw also helps to make this device practical for daily use, as battery life would not likely be a problem over the course of a day.

The AG, despite lacking an ability to generate positive net mechanical work, did demonstrate an ability to reduce amputee metabolic cost of transport for the two amputees tested. This preliminary result suggests that biarticular devices may hold the potential to improve amputee quality of life beyond that with even the most advanced monoarticular devices. These metabolic benefits may stem from a combination of a reduction in knee and hip moments and powers in the affected-side leg.

However, given the lack of statistical power, more work is needed to verify this effect on more participants. In addition, the slight increase seen in the net mechanical work from the ankle-foot prosthesis during the ACTIVE conditions could account for a nonnegligible portion of this small metabolic benefit. If, however, the metabolic improvement is confirmed for other amputees when ensuring no increase in positive work from the prosthetic ankle, it would show that metabolic improvements are possible for transtibial amputees without the need for the injection of net positive mechanical work.

It would be beneficial to perform additional experiments that more broadly explore this device. Each individual has their own gait idiosyncrasies, and a matching to other human gait solely based on height and weight has its limitations. Therefore, it is possible that, although the joint stiffness of the device was chosen via optimization, this stiffness value may not have corresponded to the optimal values for the given amputee participants. Additionally, inherent compliance in the brace itself may change the effective device stiffness from the designed specification. Future trials could more systematically vary the joint stiffness to find the metabolically optimal value and then compare it to that from simulation.

In addition, it may be desirable to reach beyond the quasi-passive model of the gastrocnemius muscle. Seeing that the biological gastrocnemius generates several joules of net work per step [43, 44], greater benefits could possibly be achieved by providing this net work to amputees. By definition, this net mechanical work cannot be produced with a quasi-passive device, and, therefore, a different mechanism would be needed to test this hypothesis. In the design of either quasi-passive or active transtibial leg prostheses, we feel biarticular gastrocnemius actuation is an important consideration.

Conflicts of Interest

The authors declare that there are no conflicts of interest regarding the publication of this article.

Acknowledgments

This work was supported by departmental funding in the MIT Media Lab.