Abstract

Due to the growing demand for assistance in rehabilitation therapies for hand movements, a robotic system is proposed to mobilize the hand fingers in flexion and extension exercises. The robotic system is composed by four, type slider-crank, mechanisms that have the ability to fit the user fingers length from the index to the little finger, through the adjustment of only one link for each mechanism. The trajectory developed by each mechanism corresponds to the natural flexoextension path of each finger. The amplitude of the rotations for metacarpophalangeal joint (MCP) and proximal interphalangeal joint (PIP) varies from 0 to 90° and the distal interphalangeal joint (DIP) varies from 0 to 60°; the joint rotations are coordinated naturally. The four R-RRT mechanisms orientation allows a 15° abduction movement for index, ring, and little fingers. The kinematic analysis of this mechanism was developed in order to assure that the displacement speed and smooth acceleration into the desired range of motion and the simulation results are presented. The reconfiguration of mechanisms covers about 95% of hand sizes of a group of Mexican adult population. Maximum trajectory tracking error is less than 3% in full range of movement and it can be compensated by the additional rotation of finger joints without injury to the user.

1. Introduction

The number of people with disabilities is increasing; thus, the demand of rehabilitation services is increasing too, due to the population growth and ageing, emerging chronic diseases, and the medical advances that preserve and extend life expectancy [1]. The World Health Organization reported “an estimated 10% of the world’s population, some 650 million people, experience some form of impairment or disability”; about 80% of people with disabilities live in developing countries. The majority are poor and experience difficulties in accessing basic health services, including rehabilitation services [1], an alternative to address this problem is the use of robotic systems in rehabilitation therapies. Robotic systems have already proven to enhance hand therapies through incorporating intensive and interactive exercises [2, 3]. Levanon confirms that “advanced technology can enrich treatment and can help patients who cannot come to the clinic regularly for treatment” [4]. “Disorders of the upper extremities specifically limit the independence of affected subjects” [5] and impairment of hand affects significantly the execution of activities of daily living (ADL). There are injuries like fractures, sprains, and dislocations that cause temporary disability and they require mobilization exercises as part of rehabilitation therapy [6]. Fasoli et al. concludes that “robotic therapy may complement other treatment approaches by reducing motor impairment in persons with moderate to severe chronic impairments” [7]. On the other hand, Carey et al. concluded “that individuals with chronic stroke receiving intensive tracking training showed improved tracking accuracy and grasp and release function, and these improvements were accompanied by brain reorganization” [8]. Thus, Kitago et al. stablish that there is a great need to develop new approaches to rehabilitation of the upper limb after stroke. Robotic therapy is a promising form of neurorehabilitation that can be delivered in higher doses than conventional therapy [9]. Additionally, rehabilitation robots also can be a platform for quantitative monitoring on the recovery process in a rehabilitation program due to the standardized experimental setup and the high repeatability of motion tasks.

Different robotic devices for upper limb rehabilitation have been developed over the past two decades to provide hand motor therapy [5]. There are different design philosophies applied to robotic therapies, determining the degrees of freedom considered and technologies used. The objective is to develop a training platform that helps patients regain hand range of motion and the ability to grasp objects, ultimately allowing the impaired hand to partake in activities of daily living [10].

In the specific case of the fingers of the hand, exoskeletons, wearable orthosis and gloves, haptic interfaces, and end-effector-based devices have been developed and evaluated in order to facilitate the rehabilitation process [3, 5]. Exoskeletons are devices with a mechanical structure that mirrors the skeletal structure of the limb; that is, each segment of the limb associated with a joint movement is attached to the corresponding segment of the device. This design allows independent, concurrent, and precise control of movements in a few limb joints. It is, however, more complex than an end-effector-based device [5]. An example of this approach is the HEXORR, Hand EXOskeleton Rehabilitation Robot [10]. This device has been designed to provide full range of motion (ROM) for all of the hand’s digits. The thumb actuator allows for variable thumb plane of motion to incorporate different degrees of extension-flexion and abduction-adduction. The finger four-bar linkage is driven by a direct current, brushless motor. The mechanisms of HEXORR only have one rotation axis for all the metacarpophalangeal joints for index to little fingers, but the rotation axes of the finger joints are not collinear. This device does not consider the distal interphalangeal joins of the fingers.

Glove devices are wearable, such as the robotic glove, which utilizes soft actuators consisting of molded elastomeric chambers with fiber reinforcements that induce specific bending, twisting, and extending trajectories under fluid pressurization. These soft actuators were mechanically programmed to match and support the range of motion of individual fingers [11]. These devices require a pneumatic or hydraulic facility, which is more complex than electric supply, especially for domestic use. The variation in hand size can be a complication for the use of these devices.

The haptic devices form another group of systems interacting with the user through the sense of touch and the mobilization of the limb. Haptic devices can be classified as either active or passive, depending on their type of actuator. An example of this approach is the “haptic knob” which is a two-degree-of-freedom robotic interface to train movements and force control of wrist and hand. The “haptic knob” uses an actuated parallelogram structure that presents two movable surfaces that are squeezed by the subject [12]. This device is oriented to perform many ADL such as grasping and manipulating objects.

The advantage of the end-effector-based systems is their simpler structure and thus less complicated control algorithms. However, it is difficult to isolate specific movements of a particular joint. The Rutgers Hand Master II is a force-feedback glove powered by pneumatic pistons positioned in the palm of the hand and provides force feedback to the thumb, index, middle, and ring fingertips [13]. The fingertips develop a linear trajectory, whose amplitude depends on the length of the pneumatic pistons. Amadeo is a commercially available device that provides endpoint control of each of the hand digits along linear fixed trajectories electric motor [14]. In this case, the fingertips develop a linear trajectory too.

The design of a reconfigurable robotic system proposed, Ro-Share, has advantages with respect to the devices mentioned. First, it is designed so that each fingertip develops a natural flexoextension trajectory considering the joint coordination of each finger kinematic chain. Each of the fingers is free to move without forcing the rotation axis alignment of its joints. Only one actuator is necessary for each mechanism that mobilizes one finger. Each mechanism can be adjusted to the finger length by the length adjustment of its crank link. The variation of hand length can be up to 16% for a male and female adult from 18 to 90 years old specific population [15]. Hence, a robotic system to guide the fingertip of fingers, index, middle, ring, and little finger, in flexion and extension exercises is proposed, which must be able to fit finger sizes through only one link length adjustment.

2. Methods

2.1. Kinematic Hand Model

The general structure of the human hand can be divided into two sets for its kinematic analysis. The first set is composed of the bones of carpus and the second of the metacarpals link and the phalanges of the five digits [16, 17]. A realistic approach, useful for flexoextension exercises of the fingers in rehabilitation therapies, considers for the fingers from the index to the little finger a kinematic chain with 3 links, which are the phalanges, and 4 degrees of freedom (DOF). For the thumb, the kinematic chain has 3 links, the metacarpal (MC), 2 phalanges, and 5 DOF. The palm of the hand is considered a fixed base. In this approach, for the index finger to the little finger, the metacarpals (MC) represent the immobile base of each kinematic chain. This is complemented by the proximal phalange (PP), medial phalange (MP), and distal phalange (DP) as links, as shown in Figure 1. The metacarpophalangeal joint (MCP) has 2 DOF, and the interphalangeal joints, proximal (PIP) and distal (DIP), have only one DOF. P represents the fingertip, the mobile end of the kinematic chain. The joins of the thumb are the carpometacarpal (TCMC) with 2 DOF, the metacarpophalangeal (TMCP) with 2 DOF, and the proximal interphalangeal joint (PIP) with one DOF.

The Denavit-Hartenberg (D-H) convention is used to define each one of the four kinematic chains [18, 19]. Table 1 presents the D-H parameters of kinematic chains, where LPP, LMP, and LDP are the nominal lengths of the proximal, medial, and distal phalanges, respectively.

Equation (1) describes the position and orientation of the fingertip [11]: where represents an array containing the position and orientation of the fingertip, and is the indicator for the fingers: index = 2, medium = 3, ring = 4, and little finger = 5, since 1 is reserved for the thumb. represents the vector used to define the reference frame at the origin of the kinematic chain of each finger relative to a fixed frame of reference at the base of the hand. are the homogeneous transformation matrices which define the geometric transformation between the origin of the kinematic chain of each finger and its tip. According to [16, 17], rotation of the finger joints is shown in Table 2. Figure 2 shows the path of flexoextension of the kinematic chain of the three phalanges of the middle finger. This path fingertip () is the initial requirement for movement of the flexoextensor mechanisms of the robotic system.

2.2. Proposed R-RRT Mechanism

The characteristics of simplicity, modularity, and low cost of construction and maintenance are desirable in devices for assistance in rehabilitation therapies [20]. The most important criterion is the patient’s safety; therefore, the range of movement of flexoextensor mechanism R-RRT is mechanically limited to avoid finger hyperextension. Accuracy in the tracking of the flexoextension path is the second aspect of importance, since continuous and gentle movements are desired from the beginning () to the end of the path (). Therefore, the position and velocity of movements should be controlled. Exercises can start with movements of low amplitude and low speed and increase both amplitude and speed up to the full range of flexion and extension. Therapies must be always supervised by professional physiotherapists to avoid sudden movements that can cause pain or further damage to the patient. In response to the above requirements, the system is composed of four four-bar type crank-slide mechanisms (R-RRT) for the mobilization of the fingers from the index to the little finger.

Figure 3 shows the corresponding mechanism design for middle finger. The DC servomotor drives point C of the slide on a linear guide through the screw, foregoing causes semicircular movements of point B of the variable length crank (VLC). Finally, point D of the coupler link develops the flexoextension path. Point D is attached to the fingertip P through a thimble secured by adhesive tape for medical use. The thimble has an unactuated joint to connect itself with link 2 (coupler) at point D; it allows the free rotation of the thimble and the natural orientation of the distal phalange. As an additional safety, a limit sensor detects when the mechanism reaches the positions of maximum flexion or extension, to indicate to an electronic controller these positions. The R-RRT, type slider-crank mechanism, is selected since the extension of the coupler link develops a very similar path presented in Figure 2. Each flexoextensor mechanism (R-RRT) develops the flexoextension path in the distal end of the extension of the coupler link (D), Figure 4(a).

The length of each finger defines the dimensions of the corresponding mechanism. The synthesis of flexoextensor mechanisms was performed from a trajectory generation method using 3 precision points based on the curves of the coupler link [21].

A major constraint is that the length of the extension of the coupling link, , must be greater than the length of the crank, , to avoid collisions between finger and mechanism. The lengths of link 2, , and the extension of the coupler link, , are defined by (2) and (3), respectively,Due to the length of the crank handle that determines the range of movement on the -axis, the path of flexoextension, Figure 2, was rotated the angle , Figure 5(a); after this, the rotated path was moved to point using a homogeneous transformation matrix H. Finally, the start and the end of the path are aligned with the axis and the position of the maximum flexion meets with point , Figure 5(b).

The rotation angle is calculated withwhere , , , and are the coordinates of the maximum flexion position () and maximum extension position () of the natural flexoextension path in the axis and . Finally, the length of the crank link is calculated withwhere and are the coordinates on the axis of the positions of maximum extension and maximum flexion of the natural flexoextension path rotated on the angle .

Once the mechanism has been synthesized, the orientation of the path is recovered by a rotation of the mechanism with the angle (equal magnitude than but with opposite direction). Afterwards, an iterative numerical method to minimize the error path was later used [22].

2.3. Reconfiguration of R-RRT Mechanism

Each patient may have a different hand size, thus requiring the reconfiguration mechanism ability to adapt to flexoextension path. Therefore, a reconfiguration scheme to modify the path of point of the coupler link is proposed in order to follow the path of flexoextension of the fingers for different sizes. For a group of Mexican population, the average hand size for adults between the ages of 18 and 90 is estimated at = 175.62 mm with a standard deviation of  mm [15]. If we consider a normal probability distribution, then, in the interval [, ], approximately 95% of the hand sizes for the Mexican population are understood. Therefore, it is estimated that hand size varies from 158.39 to 192.85 mm. The length of the crank link is half of straight length of the flexoextension rotated path on the -axis. Then, changes in the length of the crank link can modify the path amplitude of point of the mechanism. The design of a variable length crank link (VLC) is proposed in Figure 6, whose length between its external nodes and varies in proportion of length . The VLC is composed of two links connected by a passive rotational joint, point , and a screw whose end is attached to point of link 5 through a ball joint, allowing only for rotation. The other point of contact between the screw and link 4 is achieved through a nut; so that when screw turns, varies. Similar triangles formed by points and , Figure 6, have the angle , which is defined by the law of cosines aswherein and are the lengths of and lines, respectively, having the same length, .

The length is defined bywherein and are the lengths of and respectively, which have the same length, .

2.4. Kinematic Analysis of R-RRT Mechanism

In order to validate the monitoring path of flexoextension of the proposed mechanism, kinematic analysis of the synthesized mechanism is performed and the tracking error of the path is calculated. Vector loop method is used for position analysis [21]. A right-handed coordinate system is defined and a local reference frame , is located at point of fixed link 0. Each mechanism has a point , which is the point of rotation of the crank, link 1. Point is in the joint that connects the crank link to the coupler, link 2; point develops a circular constant radius equal to the crank’s length. Point , corresponding to the slide, moves on a linear axis parallel to the -axis. Finally, point , located in the extension of the coupler link, develops an elliptical path that, in the range of 180° ≤ θ ≤ 360°, is similar to the flexoextension path, Figure 2.

Vectors and correspond to links 1 and 2, respectively, Figure 4(b), so that their magnitudes are constant. Point is defined by the vector, whose magnitude varies according to angle , since it is the sum of vector and . The vector has the following components: that is parallel to the sliding axis of the slide and which is constant and corresponds to the length of offset () between the axis of the extended slide shaft and pivot of the link [21]. The most important point is , because it performs the desired path and is defined by vector, which is the result of the sum of and vectors (8). The magnitude of vector is constant, since it corresponds to the extension length of the coupler link; angle is a constant value that determines the orientation according to vector.

Position equation (8), velocity equation (9), and acceleration equation (10) of point   are calculated from the expressions:where is the length of segment , is the length of segment , and and are the angular velocity and acceleration, respectively.

3. Results

3.1. Construction Parameters

The reconfigurable robotic system for flexoextension is composed of four four-bar mechanisms and it is complemented with a variable height base on which the forearm and palm rest, Figure 7. Table 3 shows parameters of construction of four mechanisms corresponding to the index, middle, ring, and little fingers.

Figure 8 shows the behavior of the position, velocity, and acceleration of point of the mechanism of the middle finger in the - and -axes. It is observed that all the curves are smooth and continuous in the range of the path from the maximum extension () up to the maximum finger bending ().

Figure 9 illustrates the final work area of the flexoextensor reconfigurable mechanism for the case of the middle finger. The central path corresponds to the average hand size (175.62 mm) and external paths correspond to the minimum (158.39 mm) and maximum (192.85 mm) size of the Mexican adults between 18 and 90 [8]. The area between the latest two paths corresponds to the area of work expanded by reconfiguring the variable length crank link (VLC). However, after adjusting the length of the handle, it will not move until required by another patient.

3.2. Path Tracking Error

In order to determine the tracking error according to the natural course of the flexoextension path of the middle finger, the difference between the and vectors is measured. The maximum error has a value of 2.98%. Figure 10 shows the graph of error in the range of movement of the crank link.

The patient hand size is bounded in the range of 158.39 mm to 186.63 mm and is divided into eight intervals with increments of 4.30 mm. With the constant value of the length of the coupler link = 436.44 mm, length of the extension of the coupler link = 84.33 mm, and the extension of the coupler link angle β = 80° = 1.396 radians, only the length of the link is modified and the paths are shown in Figure 11. Table 4 shows the values of nine hand sizes and their maximum tracking error obtained.

3.3. Abduction Movement

In natural extension movements of the hand fingers, an abduction movement is also developed by index, ring, and little finger. When the MCP joint is flexed, the lateral ligaments are tight; it makes the abduction movements difficult or impossible [16]. The index, ring, and little fingers have major amplitude of abduction movement as 30° [Cobos]. The position of each R-RRT mechanism has the maximum flexion point as reference; in this case, the middle finger does not develop the abduction; then, middle finger mechanism orientation is parallel with the arm longitudinal axis. The mechanism orientation for index, ring, and little fingers allows an abduction movement, which begins with 0° in maximum flexion and ends with an angle γ in maximum extension. If the angle γ takes the value of the half of maximum abduction; then, γ = 15°. The result is an abduction movement of 15° for index and ring finger with reference to middle finger, and 15° of little finger with reference to ring finger. This results in comfortable movements for four fingers, index to little finger, and the metacarpophalangeal joint movement in flexoextension and abduction.

4. Conclusions

The design of the reconfigurable robotic system for flexoextension can be adapted to the size of the hand; it is the result of a compromise between functionality and practicality required for the movement of the fingers as part of rehabilitation therapies. The movement of the fingers is performed following their natural flexoextension path and prevents hyperextension, ensuring that the fingers never perform unnatural movements. The four R-RRT mechanisms orientation allows a 15° abduction movement for index, ring, and little fingers. The reconfigurability of each of the four R-RRT mechanisms allows the system to fit the hand size of each patient, allowing attention to a larger percentage of the population considered for design.

An advantage of our proposal is related to the patient comfort; that is, the patient’s hand should not carry anything, since it is based on the system. This represents an advantage over gloves or exoskeletons whose mass must be loaded by the user’s hand. The system will provide the required forces to move the fingers and allow the physiotherapist, by means of an electronic control system, to monitor and record the progress for evaluation of both the evolution of patients and rehabilitation protocols. Levanon says that “… technology is the language of the next generation and therapists must adapt the types of treatments that are able to provide their customers, and they need to be enabled to use family and significant equipment with customers in the future” [4].

Competing Interests

The authors declare that they have no competing interests.

Acknowledgments

The authors would like to acknowledge the financial support of the institutions: Instituto Politécnico Nacional, Universidad Tecnológica de Querétaro and CONACYT, also CRIQ Centro de Rehabilitación Integral de Querétaro, and the therapists for their advice and discussion about rehabilitation protocols.