Reliability of an Integrated Ultrasound and Stereophotogrammetric System for Lower Limb Anatomical Characterisation

Table 1

Individual and overall results for the three operators across n subjects (2n limbs) for all 6 variables in standing and supine positions.

VariablePositionOP001
()
OP002
()
OP003
()
Average¥
()
ICC (95% CI) §
()
value¢

Varus (+ve)/
valgus (−ve) (°)
Standing0.2 ± 3.20.1 ± 2.60.3 ± 3.20.2 ± 2.80.97 (0.86–0.99)0.61
Supine1.1 ± 2.70.4 ± 1.80.0 ± 1.60.3 ± 1.80.93 (0.71–0.99)0.63
Flexion (+ve)/
extension (−ve) (°)
Standing0.7 ± 6.24.2 ± 8.43.9 ± 8.53.6 ± 8.10.99 (0.97–0.99)0.39
Supine1.3 ± 3.41.8 ± 4.52.0 ± 4.01.7 ± 4.20.99 (0.97–0.99)0.80
Femur length (mm)Standing435 ± 31457 ± 37457 ± 35456 ± 350.99 (0.99–1.00)0.66
Supine439 ± 29454 ± 34456 ± 33455 ± 330.99 (0.99–1.00)0.49
Tibia length (mm)Standing410 ± 27426 ± 22429 ± 23428 ± 220.99 (0.98–0.99)0.55
Supine406 ± 25427 ± 21424 ± 22425 ± 210.99 (0.98–0.99)0.17
Femur torsion (°)Standing32.7 ± 10.227.2 ± 16.036.5 ± 10.331.7 ± 12.20.68 (−0.09–0.95)0.43
Supine28.6 ± 9.625.0 ± 17.926.2 ± 12.026.2 ± 13.30.95 (0.78–0.99)0.74
Tibia torsion (°)Standing30.9 ± 9.428.0 ± 6.928.2 ± 12.528.5 ± 10.40.69 (−0.61–0.96)0.17
Supine32.1 ± 8.532.4 ± 7.836.6 ± 8.633.5 ± 8.90.65 (−0.35–0.95)0.78

Mean and SD of values measured by each operator over n subjects. ¥Mean and SD across all operators for the same three subjects. §Intraclass correlation coefficient (ICC) and 95% confidence intervals (CI). ¢One-way ANOVA values.