Bilateral, Misalignment-Compensating, Full-DOF Hip Exoskeleton: Design and Kinematic Validation

Table 2

Mean, median, minimum, and maximum values of hip ROM for each of the 3 equipment states. Sagittal plane data is not available for subject 2 due to marker disappearance. Transverse plane data for subject 1 was rejected due to unauthorised movement of the torso. values are unitless and reflect the probability that the median value of the prototype (PT) and 1-DOF (1D) trial does not differ from that in the unequipped (UE) trial. As values reflect the result of a comparison with the UE state data, and they are not relevant (N.R.) for the UE state itself.

Sagittal ROM (°)Frontal ROM (°)Transverse ROM (°)
MeanMedian valueMin/maxMeanMedian valueMin/maxMeanMedian valueMin/max

S1PT113.3113.6<0.05109.8/115.647.247.6>0.139.5/51.7
UE117.4117.1N.R.115.7/119.647.045.9N.R.43.8/54.4
1D96.596.2<0.0589.2/102.016.716.6<0.0514.5/18.3
S2PT50.850.9>0.147.7/54.939.539.6<0.0538.4/40.8
UE50.150.8N.R.45.4/52.835.334.8N.R.32.6/37.4
1D23.723.9<0.0519.7/26.119.720.0<0.0514.7/24.2
S3PT115.5115.70.05 < P < 0.1113.5/117.872.761.9>0.159.5/64.047.447.5<0.0543.1/50.6
UE117.6118N.R.114.2/121.273.764.7N.R.57.5/69.545.145.9N.R.40.3/48.0
1D95.385.0<0.0581.8/112.526.926.7<0.0525.4/31.330.430.4<0.0526.0/32.6