Research Article

Solar UV-Assisted Pretreatment of River Water Samples for the Voltammetric Monitoring of Nickel and Cobalt Ultratraces

Table 5

Determination of the recovery rate for cobalt (II) in 0 hours irradiated (Original Sample), 6 and 12 hours solar UV irradiated (SoUV Sample), and 6 hours artificial UV irradiated (UV Sample).

CobaltConcentration before spiking (ppb)Spiked concentration (ppb)Concentration after spiking (ppb) Recovered concentration (ppb)Recovery rate (%)

Original0.008.003.603.6045.00
6 hrs SoUV0.002.001.301.3065.00
12 hrs SoUV0.002.001.451.4572.50
UV0.332.002.061.7386.50
Blank0.002.002.052.05102.34