We describe the design and performance of an orthogonal time-of-flight (TOF) secondary ion mass spectrometer that can be retrofitted to existing focused ion beam (FIB) instruments. In particular, a simple interface has been developed for FIB/SEM instruments from the manufacturer Tescan. Orthogonal extraction to the mass analyser obviates the need to pulse the primary ion beam and does not require the use of monoisotopic gallium to preserve mass resolution. The high-duty cycle and reasonable collection efficiency of the new instrument combined with the high spatial resolution of a gallium liquid metal ion source allow chemical observation of features smaller than 50 nm. We have also demonstrated the integration of a scanning probe microscope (SPM) operated as an atomic force microscope (AFM) within the FIB/SEM-SIMS chamber. This provides roughness information, and will also allow true three dimensional chemical images to be reconstructed from SIMS measurements.

1. Introduction and Design Philosophy

The ion microprobe or secondary ion mass spectrometer (SIMS) has existed for almost fifty years [1, 2] but although great improvements have been made in sensitivity and ease of use, the lateral resolution has not improved quite as much. Early papers reported measurements from fifty micron areas for example, [2] but submicron image resolution using gallium liquid metal ion sources (LMIS) was reported even thirty years ago [3] and the separation of features as small as 20 nm wide has been reported [4, 5], close to the limits set by sensitivity [4, 6] and lateral ion straggling [7]. However, liquid metal gallium ion beams fell out of favour for SIMS because the limited secondary ion yields did not in general allow the full lateral resolution of the small ion beam spot-size to be used, and because many applications simply demanded more sensitivity—most easily achieved by using primary ions such as Cs+ or to improve the secondary ion yield by up to two orders of magnitude. That the effective spatial resolution in SIMS is often signal limited, especially so for liquid metal gallium ion sources, is well documented, for example, [4, 6, 8]. Nonetheless, the enormous increase in interest in nanomaterials over the last decade, and the limited spatial resolution of some common chemical analysis techniques such as electron-beam-induced X-ray emission (typically restricted to a information volume of about 1 μm diameter by electron scattering) have prompted a new look at high spatial resolution SIMS [9, 10]. A further motivation to re-examine the use of a gallium LMIS for SIMS is the increasingly widespread adoption of Focused Ion Beam (FIB) instruments for sample preparation (both milling and deposition). These almost universally use gallium LMIS beams to achieve excellent spatial resolutions with a high beam brightness (corresponding to acceptable milling speeds). The coupling of a secondary ion mass spectrometer to a dedicated FIB instrument has been previously described by others for example, [5, 11, 12].

Advantages of the combination of a FIB/SEM with a SIMS include excellent sample imaging (secondary electrons from either electron or ion beams, the latter giving better channelling contrast) and positioning, and the use of the FIB to precisely expose buried regions of interest for further analysis by SIMS (and other techniques). It is far more common for FIB/SEM instruments to be designed from the start to accommodate extra analytical techniques such as EDX, EBSD, and cathodoluminescence. and these additional techniques which are complementary to SIMS can be decisive in answering analytical questions posed by complex samples. Having several types of analytical measurement possible with a single vacuum chamber is not only a great time and money saver, but also avoids uncertainty due to atmospheric contamination. The advantages of SIMS measurements over the EDX measurements more commonly found on FIB/SEM instruments are (i) better limits of detection (including fewer interferences for light elements), (ii) better spatial resolution, (iii) ability to make isotopic and molecular measurements. It is, however, more difficult in general to quantify SIMS measurements than is the case for EDX measurements on sufficiently large and thick materials.

In this paper, we describe an implementation of an FIB-SIMS instrument and show examples of measurements made with the prototype. Our focus has not been upon the best possible SIMS performance, but rather upon a pragmatic, cost-effective approach to sub-μm resolution chemical imaging that take as its starting points a standard commercial FIB instrument and a tried and tested mass analyser design. The FIB-SIMS functionality was developed as part of the European Union funded FIBLYS (see http://www.fiblys.eu/) project which aims to create a modular suite of analytical instruments which can be fitted to an optimised FIB/SEM chamber. A consequence of the decision to use a commercial FIB/SEM instrument is that the vacuum chamber is only designed to work at high vacuum (above 10−6 mbar) rather than the ultrahigh vacuum usual for secondary ion mass spectrometers. The implications of this are discussed below in the performance section.

2. The Instrument

The FIB/SEM chamber, columns, and FIB/SEM software are commercially available parts from Tescan a.s. (a Vela SEM fitted with a FIB column using a liquid metal gallium ion source). The mass spectrometer is a commercial orthogonal extraction time-of-flight instrument from Tofwerk AG (model C-TOF) with a 0.5 m nominal drift length, including the reflectron. This mass spectrometer was chosen because of its compact dimensions (see Figures 1 and 2) and because of the performance advantages and price/performance ratio of time-of-flight mass analysis compared to quadrupole mass analysis (as the samples expected to be studied by FIB-SIMS instruments are often very small and unique, the simultaneous detection of all masses in a time-of-flight instrument offers efficiency gains compared to scanning a quadrupole filter). Because of the small size of the instrument, the mass resolving power MM (FWHM) is no more than 800 at 69 Th, just sufficient to distinguish many molecular interferences from isobaric monatomic ions (Th denotes the unit Thomson for charge to mass ratio [13]). Orthogonal extraction [14], typically using a 2 μs pulse, provides a high effective duty cycle and more importantly avoids the need to pulse the primary beam (for TOF operation) or to use monoisotopic gallium (necessary to preserve mass resolution when pulsing the primary beam to define the TOF start signal), thus no changes to the FIB hardware are necessary. Secondary ions are detected in the mass spectrometer using a pair of chevron mounted microchannel plates (Photonis) and the resulting analogue signal passed through a preamplifier, an amplifier/constant fraction discriminator (RoentDek ATR19-2) and time-to-digital converter (Cronologic GmbH HPTDC8-PCI) before being passed to computer memory. Essentially the same model of mass spectrometer has been previously described in more detail in [15] for a similar application, although in the current implementation all hardware is from Tofwerk AG. Either positive or negative secondary ions can be detected, but not simultaneously. The postacceleration voltage towards the microchannel plate detector is typically 5 kV, high enough to still give detection efficiencies of about 20% for ions with masses up to 2000 Da [16].

In order to make secondary ion measurements possible, collection and transfer ion topics were designed and built to take ions from the sample to the mass analyser (mounted on a flange of the FIB/SEM chamber). A set of Einzel lenses acting as relay lenses is used to transmit ions from the collection optics to the mass analyzer; all potentials are below 200 V. A grounded shroud around the collection lens is used to improve collection efficiency and to minimize electrostatic disturbance to other instruments in the chamber. When SIMS measurements are not required, the front part of the extraction ion optics nearest the sample and pole pieces simply clips off (labelled “collection lens” in Figure 2), thus allowing the full range of sample sizes to be used by the FIB. For SIMS operation, it is advantageous to extend the ion optics so that the entrance aperture is nearer to the sample and ion beam: in the current design, the collection optics extend to within 8 mm of the sample/primary ion beam intersection. This leaves enough space for simultaneous operation of the SIMS functionality with a scanning probe microscope or gas injection system (e.g., for focused ion beam induced processing).

To collect mass spectra, software was written to (i) control the functions of the mass spectrometer, (ii) to initiate an appropriate scan series by the FIB, and (iii) to collect secondary electron images in parallel with the secondary ion mass spectra. The same software is used to view and analyse the mass spectra and to export the results in other formats (as ion images, mass spectra from selected regions, or raw data). All data are stored using the publicly available HDF5 file standard [17] which is supported by many data analysis packages such as Matlab, IDL, and Igor Pro. The TOF-SIMS software is, for performance reasons, hosted on a separate computer to that which controls the FIB/SEM. The two computers communicate via Tescan proprietary protocols using an Ethernet connection. Custom electronics and a timer card from National Instruments (model NI6602) are used to adapt hardware signals from the FIB/SEM (e.g., level shifting, logic operations, resynchronisation of clock cycles) in order to control the time-to-digital converter card and mass spectrometer extraction pulser.

During operation, the user uses the FIB/SEM to locate the point of interest and to adjust focus and magnification. Then, the desired image resolution, dwell time (mass to charge range) and number of frames are selected on the computer hosting the mass spectrometer software and the measurement is started from the latter. All triggers from the subsequent acquisition series are generated by the FIB/SEM instrument until the predetermined number of frames has been reached or the measurement is stopped manually. The FIB beam remains at each point in a frame for the defined dwell time before rapidly moving to the next point. The beam is not blanked during the move between neighbouring points, but between lines and frames it is. Although the SIMS capability has only been tested with FIB/SEM instruments from Tescan, it should be possible to fit the mass analysis hardware into instruments from other manufacturers, but the software and signalling interfaces might have to be modified. Example results and different ways of displaying the data are shown in Figures 4, 5, and 6 for a nickel finder grid.

A note on nomenclature: we refer to one complete x-y raster of the primary beam when collecting SIMS data as a “frame”. Each frame is composed of data from an evenly spaced Cartesian grid of points projected onto the sample surface. From each of these points, one mass spectrum is obtained from one extraction cycle of the TOFMS. Because dynamic SIMS implies sputtering, we may consider each measurement point as having some depth, and thus being a volume element or voxel. These are referred to as “raw voxels”. In practice, because the number of ions detected per raw voxel is rather low, these raw voxels are binned into “effective voxels” for visualization although the raw data is kept. It should be noted that SIMS measurements alone do not provide depth information, thus the use of “voxel” might be argued with. The meaning of “depth profile” and “chemical image” is illustrated in Figure 3.

3. Performance

Experience with this design of mass spectrometer suggests that the dynamic range of the instrument is limited by internal ion scattering to about eight decades (implying 10 ppb detection limits if all secondary ions were produced with the same useful yield). In practice, for most samples integration times sufficient to make 10 ppb peaks clear above the scattered ion background are impractical using primary ion currents suited to submicron resolution imaging (<200 pA). Figure 6 shows in excess of six decades of dynamic range, and Figure 7 shows a calibration curve for boron in silicon from which a limit of detection of 5.5 ppm is inferred. The limit of detection is a strong function of the secondary ion yield, so it will vary for different secondary ions and matrices.

The useful ion yield was calculated for aluminium secondary ions from a distributed Bragg reflector (DBR) structure consisting of alternating AlGaAs and GaAs layers of known thickness (see Figure 12), and a figure of 4 × 10−6 detected ions per sputtered atom was obtained using a primary ion energy of 20 keV. We are not aware of a comparable useful yield measurement in the literature for the same material and primary ion beam. It is well known that secondary ion yields vary enormously with the ion and matrix (as well as primary ion element, energy, and incident angle) and surface quality. Useful ion yields are further dependent upon the tuning of the mass analyser and extraction optics (it is for example common practice to use an energy filter to suppress polyatomic ions despite the resulting reduction in useful ion yield for the analyte ions); a comparison to the (sparse) literature is, therefore, somewhat difficult. Reported values for useful ion yields for gallium ion SIMS include 1.4 × 10−2 for Si+ from silicon oxide, 6 × 10−5 for Si+ from silicon, 1.93 × 10−3 for In+ from indium phosphide and 7.8 × 10−6 for P+ from indium phosphide [18, 19]. Bearing in mind that aluminium ions seem to be detected more sensitively than most on our instrument and that the chamber pressure is always above 4 × 10−6 mbar (and so some signal enhancement due to oxygen possible) and there is a suggestion that the observed useful ion yield could be improved. We did not use silicon or silicon dioxide to calculate a useful ion yield, because it was not possible to obtain a stable signal under normal conditions at the pressure used in the instrument described here. The implications of high vacuum operation of a SIMS instrument (instead of ultrahigh vacuum) are discussed below.

Chemical images in which lines spaced by less than 100 nm are separated have been obtained (see Figures 9 and 13), and we are certain that this figure can be improved upon in future by better focussing of the FIB beam. The depth resolution can be, as expected, much better than 20 nm; see Figure 15.

Mass Discrimination
The gallium isotope ratio from a GaAs target (so the gallium signal consists of secondary ions from both the sample and from implanted primary ions) was measured as 1.46 ± 0.02, the literature value is 1.507 ± 0.0004 [20]. Mass discrimination is therefore of the order of 15‰  Da for this example. Similar magnitudes for the mass discrimination have been observed for iron and nickel from metal alloys.

Mass Resolution and Abundance Sensitivity
The measurements reported here were obtained using a mass resolving power of MM (FWHM) of 400 Th/Th at mass to charge ratio 69 Th. The abundance sensitivity was 2.5 × 10−5.

3.1. Effect of Chamber Pressure

Commercial FIB/SEM instruments are mostly designed to work at high vacuum rather than ultrahigh vacuum, and this has several consequences when performing secondary ion measurements in such an instrument.(i)Because secondary ion yields depend upon the partial pressure of for example, oxygen (e.g., [21]), there will be a change in sensitivity (drift) as the instrument pumps down after opening the chamber, for example, to change samples. The presence of oxygen in conjunction with ion beam sputtering can complicate the appearance of oxide interfaces [22].(ii)Even at the lowest pressure the chamber can attain (>1 × 10−6 mbar), there will be a sufficient partial pressure of chemically active gases (e.g., water and oxygen) to affect the secondary ion yield for some materials [23, 24]. This may improve sensitivity, but complicates comparison of useful yields to published values. See Figure 8 for an example of this effect.(iii)The absorption of chamber gases onto the sample surface contributes a background to the secondary ion mass spectrum, and causes a nonlinear dependence of the SIMS signal upon the primary beam current and scan area [25, 26].(iv)The presence of active gases changes the extent to which surfaces roughen under ion bombardment.

All of these effects may be minimised (but not avoided completely) by venting with an inert gas or better by using a load-lock chamber. Normalisation to the gallium ion signal can sometimes be used to correct for oxidation effects on SIMS yields [27]. In general, however, for good reproducibility it is necessary to record (and to control) the pressure in the chamber when performing SIMS measurements at high vacuum. For samples which are sensitive to the residual gases in the chamber (e.g., silicon), we suggest deliberately adding a known partial pressure of for example, oxygen in order to improve stability. Oxygen flooding can however have side-effects such as distortion of depth profiles [24].

4. Applications

To illustrate the performance of our first FIB-SIMS prototype, we present here some example applications. When an image is captioned as a depth profile, it means that the data has been integrated from a region of each x-y plane (frame) and is plotted as a function of frame number (equivalent to depth or z-direction). When an image is captioned as an “image”, it means that the x-y (frame) data is shown, and that the data has been integrated along the z/depth direction. See also Figure 3. For all of the data shown here, the primary ion beam was incident normal to the mean sample surface. Depth profiles use only the centre part of each frame in order to minimise crater edge effects (this conservative approach throws away 75% of the signal).

4.1. Spatial Resolution Standard BAM L200

Figure 9 shows an aluminium secondary ion image from an unmounted, unpolished chip of the BAM L200 resolution test reference material. BAM define the spatial resolution to be that of the smallest period grating that can be distinguished, in this case P7 with period 67.5 nm (the widths and separation of the aluminium rich stripes in P7 are 34 nm). The ion beam FWHM is calculated to be about 50 nm (see below). Noteworthy is that the chemical image was achieved from a rather rough surface—the existence of surface topography (scratches from sawing the sample) did not distort the chemical image. It was also possible to detect (but not resolve) the single aluminium rich stripes W8 (38 nm), W5 (19. nm) W10 (14.2 nm) and W9 (3.6 nm). In a round-robin study of SIMS resolution [10] only ten out of sixteen laboratories could detect a 5 nm wide aluminium rich stripe in the similar BAM L002 standard, and the smallest calculated beam FWHM was 69 nm.

Figure 10 shows the corresponding indium ion image. The signal here is nosier both because indium is less sensitively detected than aluminium from this material and because the indium-rich stripes are very thin (<5 nm, much smaller than a single pixel in this image).

The lateral resolution of an image can be defined in various ways, as discussed by [28] and will, in general, depend also upon the signal to noise ratio of the image. The ISO definition for surface chemical analysis makes reference to the distance between the 12% and 88% intensity points in a line scan across a well-defined step function [29]. For a Gaussian beam, a good approximation to gallium LMIS beam spot distributions [10], this distance corresponds to the full width at half-maximum (FWHM). Note that some FIB manufacturers prefer to use the 20% and 80% intensity points. Figure 11 shows the result of fitting the complementary error function (the convolution of a Gaussian with a step function) to part of the aluminium profile shown in Figure 9. The aluminium rich stripe in P4 has a width of 97 nm; just large enough compared to the spot size to regard the profile as a measured edge spread function. A reasonable fit is obtained suggesting that the beam profile is well approximated by a Gaussian distribution. Using either the ISO definition or a deconvolution of the data gives a FWHM for the primary beam of about 50 nm (the 20%–80% definition gives a spot size of about 40 nm). With experience, we expect further increases in performance, as according to the manufacturer it should be possible to obtain this beam size at much higher currents than we used (the specification is for 50 pA into a 20 nm spot, or 500 pA into a 50 nm spot).

The above analysis assumes that the resolution is determined by the primary beam size and not by any other physical processes. We can neglect lateral straggling of primary ions as this is of the order of a few nanometers [7], possibly up to 20 nm [9]. Although differential sputtering and the resulting increase in rugosity of the surface might contribute to a larger effective spot size, because secondary ion yields depend on surface roughness and the surface-primary beam angle, we do not see any change of lateral resolution with depth.

4.2. Semiconductor Laser

The performance achieved with the BAM L200 standard is particularly relevant to applications involving semiconductor or photonic devices fabricated in gallium arsenide. Figures 12 to 14 show depth profiles and chemical images of a Vertical Cavity Surface Emitting Laser constructed from an AlGaAs/GaAs Distributed Bragg Reflector (DBR) fused to an AlGaInAs/InP gain wafer. The device has been previously described in more detail in [30] and was the first example of a wafer-fused semiconductor disk laser operating at the 1.3 μm wavelength. The specimen described here is denoted VECSEL 3/1. Figure 12 shows a positive secondary ion depth profile through the device. On the right hand side is part of the AlGaAs/GaAs DBR structure (35 repeats of a 100 nm layer of GaAs and a 100 nm layer of Al0.9Ga0.1As). On the left-hand side (left of frame 70 on the x-axis) are the structures in the AlGaInAs/InP wafer. The expected structure is shown schematically in Figure 13. Although much of the compositional structure of the device is visible, the structure within the five quantum well regions was not resolved. (Each of these 44 nm thick units consists of three 10 nm thick Al0.28Ga0.26In0.46As layers separated by two 7 nm thick layers of Al0.14Ga0.18In0.68As; each unit is then spaced apart by InP layers). The depth resolution in this profile is apparently of the order of 20 nm. As reported previously for a SIMS depth profile of a similar material [31] using different experimental conditions, we see a decrease in the modulation depth of the repeated layers with depth (i.e., the peak to valley difference of the aluminium signal) which was shown to be due to roughening of the surface (ripple formation [31]). The effect is less severe for the conditions reported here, consistent with the lower extent of roughening as measured by scanning probe microscopy (see section below).

Figure 13 shows the same structure, but in a different experiment viewed from the side. The 100 nm thick aluminium stripes in the DBR are clearly separated from each other, but do not have flat tops, and the modulation depth is only about 80%, so the resolution in this image is a little better than 100 nm.

That the FIB-SIMS cannot only visualise expected chemical structures, but also reveal new information, is shown in Figure 14. This overlay of depth profiles obtained from both positive and negative secondary ions shows an unexpected small peak in the oxygen ion signal at frame 60—this is very close to the wafer interface and suggests that a surface oxide layer may have been incorporated during the wafer fusion process.

4.3. Atomic Layer Deposition Fabricated Multilayers

Atomic layer deposition (ALD) is a coating technique using self-limiting surface reactions of alternately supplied volatile precursor molecules to make highly conformal films with excellent control of thickness [32], almost to single atomic layers. Conformal ALD coatings have been used for example to coat nanopillars for novel applications in photovoltaic materials and nanomechanics, as well as to make transparent conductive oxide films. Zinc oxide has potential as a semiconductor material for optoelectronics, and aluminium-doped zinc oxide can be used as a thin transparent conducting oxide (TCO) film. The FIB-SIMS was used to investigate ALD multilayers of zinc oxide and aluminium oxide that were prepared as part of a study into their photoelectric properties, and in particular to study some fundamental characteristics of the ALD produced materials.

The films were produced at EMPA in a home-made reactor using dimethyl zinc (95% purity from STREM Chemicals, Inc.), trimethyl aluminium (97% purity from Sigma-Aldrich Co.) and water vapour as reagents, and argon as a carrier gas. The substrate was (111) silicon carefully cleaned to remove organic contaminants from the surface. The deposition was carried out at 65°C with the following cycle times: 0.1 s pulse, 20 s exposure, and 45 s purge. A 20 nm thickness of Al2O3 required 154 cycles and a 20 nm thickness of ZnO required 100 cycles. Figure 15 shows a FIB-SIMS depth profile through five pairs of nominally 20 nm zinc oxide and 20 nm aluminium oxide layers on a silicon substrate. All of the layers are clearly seen in the depth profile and the aluminium signal shows a “top-hat” profile suggesting that the depth resolution in aluminium oxide is much better than 20 nm. However, the modulation depth of the aluminium signal is poor and the zinc signal does not have a “top-hat” profile. Figure 16 shows that the zinc oxide layers are actually thinner than the aluminium oxide layers, furthermore we expect the zinc oxide layers to be sputtered faster. As a result, the zinc oxide layers although clearly visible are not thick enough to give a “top-hat” profile. It is not clear why the zinc signal from the uppermost layer (and to a lesser extent the aluminium signal too) is so much stronger than for subsequent layers. We suspect that this is a result of adsorbed atmospheric constituents (molecular oxygen, water) in the ALD material. The signature is probably not simply due to a build up to the equilibrium concentration of implanted gallium, because it does not repeat at each subsequent interface. (For 20 keV primary ion energy, we do observe repeated structure in the SIMS depth profile within each aluminium oxide layer; this structure is believed to be a measurement artefact. SIMS measurements at interfaces can be complicated because sputtering can cause chemical changes (e.g., reduction) in the sample material at the same time that the implanted ion dose and damage approach a steady state [22]). The high signal for the mass to charge ratio 28 Th between the last aluminium oxide layer and the silicon substrate is due to chemistry at the substrate interface resulting in (i) either a stronger AlH+ ion signal (which interferes strongly with 28Si+, and in practice dominates in this sample) or (ii) a stronger 28Si+ signal due, for example, to the presence of a thin layer of silicon oxide. Isotope ratios suggest that the latter explanation (enhancement of silicon secondary ion signal at the interface) is more likely. An unexpected observation was that of a weak boron signal (see Figure 17), correlating with the zinc oxide layers. The presence of boron in ALD fabricated zinc oxide layers was subsequently confirmed by glow discharge optical emission spectroscopy (GD-OES), and was found to be due to an impurity in the zinc precursor.

Limiting Depth Resolution of SIMS
Secondary ions come from very near the surface (<0.5 nm) and the depth resolution of a secondary ion depth profile, given a sufficient signal to noise ratio, is in principle only limited by ion-induced mixing (usually of the order of a few nanometers but can be more depending on primary ion energy and angle and sample composition) and ion-induced roughening, in common with other techniques which use an ion beam to sputter away material in order to obtain a depth profile (e.g., XPS, Auger). Changes in the composition of the sample usually affect secondary ion yields, and differential sputtering (in the chemical sense rather than the topographical sense) will have an effect on depth profiles at interfaces [33]. The processes that limit SIMS depth resolution are discussed in detail in [34]. It is worth noting that we have no evidence for any background due to sputtered material being resputtered and ionised from the SIMS ion optics (i.e., no memory effect is detectable). Redeposition of material from the crater walls onto the crater floor can affect depth profiles [34, 35], the extent to which this matters will depend upon the chemical structure of the sample, the size of the crater and the size and shape of the primary ion beam. With appropriate mathematical models, monolayer thicknesses can be reconstructed from SIMS depth profiles [36]. Note that the mixing-roughness-information depth (MRI) model proposed by Hofmann [36] ideally requires knowledge of the original surface topography, any interface topography, and of the sputtering induced roughness. A scanning probe microscope can provide at least two of these.

4.4. SPM-SIMS Integration

Because dynamic secondary ion measurements imply significant sputtering, in general ions originating from different depths will be measured at different times and thus the recorded dataset is in some sense three dimensional [37, 38]. However, for heterogeneous samples there is no simple relationship between sputtering time or ion dose and the depth. This is because different materials and different crystal orientations will in general have different sputtering rates [39]. Furthermore, if the sample does not initially have a flat polished surface, then there will always be ambiguity about the spatial relationship between observed features as sputtering proceeds. Even for polished samples where the matrix is homogenous, surface roughening and swelling [40] due to the ion beam can distort the relationship between sputtering time or dose and the depth being sampled. Artefacts in the SIMS image may also arise due to imperfectly normal crater walls, because of edge effects when topography develops on the surface being measured and due to redeposition of sputtered atoms [35], and because local topography affects useful secondary ion yields [41]. For all of these reasons it is highly desirable to have available during sputtering a measurement of the surface profile relative to the original surface before sputtering commenced. The best way of doing this is to use a scanning probe microscope to measure the surface (other options such as stereo-pair electron beam imaging do not have sufficient depth resolution or accuracy; laser interferometry has been used to obtain sputtering rates [42] but gave no information on roughening or differential sputtering within a crater and can be difficult to interpret [43]). The use of an ex situ SPM to improve SIMS 3D images has been previously described for example, by [31, 39, 44]. In situ measurements are highly desirable to increase the speed of measurements and to avoid concerns about contamination or reactions at the surface. This is however not commonplace because (i) few scanning probe microscopes are intended to work under vacuum, (ii) fewer still can easily be integrated with an electron microscope and focused ion beam (limited space, possible interactions), (iii) scanning probe microscopes are expensive, and (iv) combining SPM measurements with SIMS measurements can enormously increase the required measurement time. However, If an SPM is integrated with a SIMS measurement, then it opens up the possibility of not only sensing topography but also other properties of the surface (e.g., contact potential, magnetisation, and friction) complementary to the composition measured by SIMS. Because we had prior experience with SPM/SEM integration, it was natural to do some “proof of concept” experiments to demonstrate SPM/FIB/SEM-SIMS integration within a single vacuum chamber. We are unaware of any prior publications that show in situ SPM measurements of SIMS craters.

The in situ scanning probe microscope was developed together with SPECS Zurich GmbH. It utilised SmarActGmbH positioners for coarse control of the tip and closed-loop Nanodrive piezo stages from Mad City Labs Inc. for z tip control and x-y fine positioning (x-y scanning was done by moving the sample rather than the probe tip). Frequency modulation atomic force microscopy (AFM) was performed with an Akiyama probe from Nanoworld AG [45]. A Nanonis SPM control system (base package and oscillation controller) was used for the probe amplitude and phase feedback loops, and to monitor the frequency shift of the probe. The geometry and self-sensing properties of the Akiyama probe are particularly well suited to a compact integration within the vacuum chamber close to the intersection of the charged particle beams with the sample. When not in use, the SPM probe tip was retracted from the measurement region using the coarse positioner, allowing free access of the charged particle beams to the sample, and unhindered collection of secondary ions and electrons. The SPM was mounted directly onto the sample stage of the FIB/SEM as shown in Figure 18. Although the Akiyama probe signal can be transduced in different ways, we illustrate here only the calculated Z displacement of the tip due to the surface being measured. The free software WSXM [46] was used to visualise and analyse the SPM datasets. An SPM scan in both directions over a 5 × 5 micron area typically took about 20 minutes (this obviously depends on the roughness, whether the crater edge is part of the scan, and the desired resolutions). The time to automatically withdraw the SPM tip so that FIB scanning can continue is about 15 seconds; more critical to combining the use of FIB-SIMS and SPM is the successful automatic coarse repositioning of the SPM tip.

The semiconductor laser sample described above was used as a test specimen. We were particularly interested to find out if surface roughening could have been affecting the depth resolution of the SIMS measurements. AFM scans were performed before, during, and after the SIMS measurement. (“during” meaning that the FIB beam was briefly turned off, and the AFM tip moved into position, before reverting to SIMS mode and continuing to sputter).

Figures 19 and 20 illustrate that the crater floor was fairly smooth (RMS roughness 4.9 nm at a depth of 750 nm after 200 frames), so ion-induced roughening cannot have been significantly limiting the extent to which the aluminium-rich stripes in the DBR were resolved although this degree of roughening will certainly have been detrimental to the resolution of the 7 and 10 nm deep features in the laser gain medium (quantum well structures). The crater walls, as imaged by the AFM, are not perfectly perpendicular to the sample surface, but there is little evidence for significant redeposition of material around the crater. It seems likely that the main reason for the unexpectedly poor modulation and resolution of the 100 nm thick aluminium structures is a combination of sputtering of redeposited material from within the crater and primary beam halo effects.

5. Conclusions

We have shown that it is possible to adapt a commercial FIB/SEM instrument for SIMS operation by adding a time-of-flight mass analyser without any detrimental effect upon normal operation. A depth resolution of <20 nm and a lateral resolution of <100 nm (see Figure 9) have been demonstrated (both for normally incident ions), and we expect to improve upon both of these figures in the near future. We have also demonstrated that the additional integration of a scanning probe microscope is possible, allowing true depth and roughness information to be obtained in series with SIMS measurements.

Acknowledgments

The authors gratefully acknowledge the financial support of the European Union for part of this work (FP7 Project FIBLYS, Grant Agreement no. CP-TP 214042-2) and funding from the Swiss Commission for Technology and Innovation with SPECS Zürich supporting the SEM/SPM development, (CTI-Project no. 9208.1 PFNM-NM). The authors would also like to thank Gerhard Bürki of EMPA for technical assistance with the FIB/SEM and Dr V Iakovlev and Dr. A. Sirbu of EPFL for providing the VCSEL sample. For “Advances in Materials Science and Engineering” Dr. A. G. Balogh and an anonymous reviewer helped to improve the quality of this paper.