Advances in Orthopedics
Volume 2013, Article ID 347358, 5 pages
http://dx.doi.org/10.1155/2013/347358
Clinical Study

Cement Removal from the Femur Using the ROBODOC System in Revision Total Hip Arthroplasty

1Center of Arthroplasty, Kyowakai Hospital, 1-24-1 Kishibekita, Suita, Osaka 564-0001, Japan
2Department of Orthopaedic Surgery, Osaka University Graduate School of Medicine, 2-2 Yamadaoka, Suita, Osaka 565-0871, Japan
3Department of Orthopaedic Medical Engineering, Osaka University Graduate School of Medicine, 2-2 Yamadaoka, Suita, Osaka 565-0871, Japan

Received 15 September 2012; Accepted 9 September 2013

Academic Editor: Michael A. Conditt

Copyright © 2013 Mitsuyoshi Yamamura et al. This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

Abstract

Introduction. The perforation and fracture of the femur during the removal of bone cement in revision total hip arthroplasty (THA) are serious complications. The ROBODOC system has been designed to selectively remove bone cement from the femoral canal, but results have not been reported yet. The purpose of our study was to evaluate the clinical and radiographic results of revision THA using the ROBODOC system for cement removal. Materials and Methods. The subjects comprised 19 patients who underwent revision THA using the ROBODOC system. The minimum duration of follow-up was 76 months (median, 109 months; range, 76–150 months). The extent of remaining bone cement on postoperative radiography, timing of weight bearing, and the complications were evaluated. Results. The mean Merle d’Aubigne and Postel score increased from 10 points preoperatively to 14 points by final follow-up. Bone cement was completely removed in all cases. Full weight bearing was possible within 1 week after surgery in 9 of the 19 cases and within 2 months in all remaining cases. No instances of perforation or fracture of the femur were encountered. Conclusions. Bone cement could be safely removed using the ROBODOC system, and no serious complications occurred. Full weight bearing was achieved early in the postoperative course because of circumferential preservation of the femoral cortex.