Abstract

Purpose. To evaluate to which extent individual Zernike terms can be corrected. Methods. Ablation time and fidelity was analysed using different fluence levels (range 90–2000 mJ/) and aspheric ablation profiles. With optimal parameters, the extent to which individual Zernike modes can be corrected was evaluated. Results. The range 188–565 mJ/ resulted as optimum fluence level with an optimum proportion range 50%–90% for high fluence. With optimal parameters, it corresponds to 2.4 s/D at 6 mm OZ, with fidelity variance of 53 m RMS, and average ablation error of 0.5 m for each location. Ablation simulation of coma Z[3,1] showed 98,4% accuracy and 98% fit quality; trefoil Z[3,3], 99,9% accuracy and 98% fit quality; spherical aberration Z[4,0], 96,6% accuracy and 97% fit quality; secondary astigmatism Z[4,2], 97,9% accuracy and 98% fit quality. Real ablation on a flat plate of PMMA of coma Z[3,1] showed 96,7% accuracy and 96% fit quality; trefoil Z[3,3], 97,1% accuracy and 96% fit quality; spherical aberration Z[4,0], with 93,9% accuracy and 90% fit quality; secondary astigmatism Z[4,2], with 96,0% accuracy and 96% fit quality. Conclusions. Ablation of aspherical and customised shapes based upon Zernike polynomials up to the the 8th order seems accurate using the dual fluence concept implemented at the AMARIS platform.

1. Introduction

With the introduction of the laser technologies for refractive surgery, the change of the corneal curvature to compensate in a controlled manner for refractive errors of the eye [1] is more accurate than ever. The procedure is nowadays a successful technique, due to its submicrometric precision and the high predictability and repeatability of corneal ablation accompanied by minimal side effects. Standard ablation profiles based on the removal of convex-concave tissue lenticules with spherocylindrical surfaces proved to be effective in the compensation of primary refractive errors. However, the quality of vision deteriorated significantly, especially under mesopic and low-contrast conditions [2].

With the introduction of wavefront analysis, it was proved that the conventional refractive LASER techniques were far from ideal, by measuring the aberrations induced by conventional algorithms and the aberrations induced by the LASIK flap cut itself.

With the LASIK (Laser in Situ Keratomileusis [3]) treatment, we have an accepted method to correct refractive errors such myopia [4], hyperopia [5], and astigmatism [6]. One of the most significant side effects in myopic LASIK, is the induction of spherical aberration [7], which causes halos and reduced contrast sensitivity [2]. However, the different laser platforms are always introducing new concepts and optimising their ablation profiles.

Since Laser refractive surgery was introduced, the technology rapidly improved. With the beginning of photo-ablation, the goal was to achieve predictable and stable results for myopic, hyperopic, and astigmatic corrections. Today’s technology is far more advanced since sophisticated diagnostic instruments, such as aberrometers and topography systems, offer the challenge of improving the postoperative results in terms of visual acuity and night vision. At the same time, the better knowledge and understanding on refractive surgery by potential patients upgrades the required standard outcomes. Making more challenge finding new approaches towards the close-to-zero aberrations target results in several senses: (a) finding the sources of the induced aberrations due to laser refractive surgery, (b) developing “free-of-aberrations” ablation profiles, (c) developing ablation profiles to compensate the natural aberrations of any single eye in order to get a close-to-zero aberrations result.

To eliminate already existing aberrations, the so-called “customized” treatments were developed. Customisation of the ablation is possible either using wavefront measurements of the whole eye [8] (obtained, e.g., by Hartman-Shack wavefront sensors) or by using corneal topography-derived wavefront analyses [9, 10]. Topographic-guided [11], Wavefront-driven [12], Wavefront-optimized [13], Asphericity preserving, and Q-factor profiles [14] have all been put forward as solutions. Nevertheless, considerations such as treatment duration, tissue removal [15], tissue remodelling, and overall postoperative outcomes have made it difficult to establish a universal optimal profile.

Therefore, the topic “ablation resolution in laser corneal refractive surgery” is still worth to be analysed and considered, because its clinical implications are not yet deeply explored.

The real impact of ablation resolution in laser corneal refractive surgery is still discussed in a controversial way. The aim of this work is to provide a simple and understandable theoretical frame explaining a possible method of ablation resolution optimisation.

Most of the systems available for laser refractive surgery include possibilities to customise the ablation, either based on topographical elevation or on corneal or ocular wavefront aberration.

2. Materials

To evaluate the technical capabilities to correct individual Zernike modes, and the extent to which individual Zernike modes can be corrected, the CAM software was used to plan the ablations, which were first simulated and then ablated onto flat PMMA plates with an AMARIS excimer laser (SCHWIND eye-tech-solutions GmbH, Kleinhostheim, Germany).

The AMARIS laser system works at a repetition rate of 500 Hz, produces a spot size of 0.54 mm (full width at half maximum (FWHM)) with a superGaussian ablative spot profile [16, 17]. High-speed eye-tracking with 1050 Hz acquisition rate is accomplished with 3-ms latency period [18]. The system delivers aspheric wavefront-customised profiles and including some optimisations: The aspheric profiles go beyond the Munnerlyn proposed profiles, and add some aspheric characteristics to balance the induction of spherical aberration (prolateness optimisation).

These particular case of aspheric profiles compensate for the aberrations induction observed with other types of profile definitions [19], some of those sources of aberrations are those ones related to the loss of efficiency of the laser ablation for nonnormal incidence [2022].

Optimisation consisted to take into account the loss of efficiency at the periphery of the cornea in relation to the centre as there is a tangential effect of the spot in relation to the curvature of the cornea (Keratometry (K-reading)). The software provides K-reading compensation, which considers the change in spot geometry and reflection losses of ablation efficiency.

An optical zone of 6.50 mm in diameter was planned, a variable transition zone, to smooth the ablated area towards the nontreated cornea, was provided automatically by the software in relation to the refraction to be corrected.

Real ablative spot shape (volume) is considered through a self-constructing algorithm. In addition, there is a randomised flying-spot ablation pattern, and controls the local repetition rates to minimise the thermal load of the treatment [23] (smooth ablation, no risk of thermal damage). Therefore, the ablated surface after aspheric wavefront-customised profiles is very smooth, so that there are some benefits in higher order aberrations.

3. Method

Laser corneal refractive surgery is based on the use of a Laser (typically an excimer one) to change the corneal curvature to compensate for refractive errors of the eye. It has become the most successful technique, mainly due to the submicron precision and the high repeatability of the ablation of the cornea accompanied by minimal side effects. Laser refractive surgery is based upon the sequential delivery of a multiplicity of laser pulses each one removing (ablating) a small amount of corneal tissue.

From the blowoff model (derived form the Beer-Lambert’s law), the real energy density absorbed at that point determines the ablation depth as where is the actual depth per pulse at the location ; is the radiant exposure of pulse at the location ; is the reflectivity at the location ; is the corneal threshold; is the corneal absorption coefficient.

In general, For different beam profiles we get different spot profiles as depicted in Figures 1 and 2.

For different radiant exposures with the same beam profiles, we get different spot profiles as displayed in Figure 3.

The problem of the spot profile and the radiant exposure of the system relies on the sequential delivery of a multiplicity of laser pulses each one ablating locally a small amount of corneal tissue, being the global process an integral effect. The higher the spot profile is, the higher the ablated volume per pulse limiting the resolution of the treatment.

There are several ways to avoid that problem.(i)Reducing the radiant exposure improving the vertical resolution of the treatment.(ii)Reducing the spot diameter improving the horizontal resolution of the treatment.

The problem of both alternatives is that they need extra time for the ablation procedure, which may produce another inconveniences.

The gained ablation volume has to be applied onto the cornea by thousands of single laser shots at different but partly repeated corneal positions, because the ablated volume of a single spot is much smaller than the ablation volume.

We have introduced as well some innovations concerning ablation shot file (sequences of pulses needed to carry out a refractive procedure) generation, in order to optimally remove the tissue corresponding to these state-of-the-art treatments, generating the sequence of laser shot coordinates in a way that(i)guarantees a high fidelity reproduction of the given ablation volume line shape and(ii)avoids vacancies and roughness of the cornea.

In this context, two opposed requirements define the fluence level.(i)A short ablation time (favouring high fluence levels).(ii)A high-fidelity ablation (favouring low fluence levels).

We have analysed ablation times and fidelity using fixed fluence levels ranging from 37 pl spot volume to 440 pl spot volume (corresponding to 90 mJ/cm2 to 2000 mJ/cm2 energy density, using as ablation volume the one corresponding to an aberration-free correction of 5.00 D–3.50 D 1 at 6.50 mm OZ, 8.21 mm TAZ.

A second simulation was prepared upon the basis of a dual fluence level concept, using a variable rate from only low-fluence spots to only high-fluence spots using the same ablation volume for the correction of 5.00 D–3.50 D 1 at 6.50 mm OZ, 8.21 mm TAZ.

With the optimal parameters we have evaluated the extent to which individual Zernike modes can be corrected, first simulated and then ablated onto flat PMMA plates.

4. Results

We have analysed ablation times and fidelity using fixed fluence levels ranging from 37 pl spot volume to 440 pl spot volume (corresponding to 90 mJ/cm2 to 2000 mJ/cm2 energy density, using as ablation volume the one corresponding to an aberration-free correction of 5.00 D–3.50 D 1 at 6.50 mm OZ, 8.21 mm TAZ (Figure 4).

We have obtained as optimum fluence level a variable range from 105 pl spot volume to 240 pl spot volume, corresponding to 188 mJ/cm2 to 565 mJ/cm2 energy density with equivalent ablation time ranges from 60 seconds to 26 seconds, corresponding to 4.0 s/D to 1.7 s/D at 6 mm OZ and equivalent fidelity variance from 50 m RMS to 60 m RMS and average ablation error from 0.475 to 0.575 m for each location (Table 1).

A second simulation was prepared upon the basis of a dual fluence level concept, using a variable rate from only low-fluence spots to only high-fluence spots using the same ablation volume for the correction of 5.00 D–3.50 D 1 at 6.50 mm OZ, 8.21 mm TAZ (Figure 5).

We have obtained an optimum proportion range variable from 50% high fluence to 90% high fluence, with equivalent ablation time range from 45 seconds to 33 seconds, corresponding to 3.0 s/D to 2.2 s/D at 6 mm OZ, with equivalent fidelity variance from 52 m RMS to 55 m RMS, and average ablation error from 0.490 to 0.525 m for each location (Table 2).

With the optimal parameters we have simulated the extent to which individual Zernike modes can be corrected (Figures 6, 7, 8, and 9, Table 3).

With the same parameters, we have evaluated the extent to which individual Zernike modes can be corrected, by ablating onto flat PMMA plates (Figures 10, 11, 12, and 13).

5. Discussion

We have evaluated to which extent individual Zernike terms can be corrected, by analysing ablation times and fidelity using different fluence levels (range 90–2000 mJ/cm2), and aspheric ablation profiles, as well as, using a dual fluence level concept (variable rate from only low fluence to only high fluence pulses). With the optimal parameters, the extent to which individual Zernike modes can be corrected was simulated and ablated onto PMMA with a laser.

Huang and Arif [16] investigated the effect of laser spot size on the outcome of aberration correction with scanning laser corneal ablation using numerical simulation of ablation outcome of correction of wavefront aberrations of Zernike modes from second to eighth order. They modeled gaussian and top-hat beams from 0.6 to 2.0 mm full-width-half-maximum diameters, evaluated the fractional correction and secondary aberration (distortion), and used a distortion/correction ratio of less than 0.5 as a cutoff for adequate performance. The found that a 2 mm or smaller beam is adequate for spherocylindrical correction (Zernike second order), a 1 mm or smaller beam is adequate for correction of up to fourth order Zernike modes, and a 0.6 mm or smaller beam is adequate for correction of up to sixth order Zernike modes.

Guirao et al. [17] calculated that the success of a customized laser surgery attempting to correct higher order aberrations depends on using a laser beam that is small enough to produce fine ablation profiles needed to correct higher order aberrations. Simulating more than 100 theoretical customized ablations performed with beams of 0.5, 1.0, 1.5, and 2.0 mm in diameter, they calculated the residual aberrations remaining in the eye and estimated the modulation transfer function (MTF) from the residual aberrations. They found that the laser beam acts like a spatial filter, smoothing the finest features in the ablation profile and that the quality of the correction declines steadily when the beam size increases. A beam of 2 mm was capable of correcting defocus and astigmatism. Beam diameters of 1 mm or less may effectively correct aberrations up to fifth order.

Pettit [24] claimed that the LADARVision system using a small fixed diameter excimer laser beam providing a consistent ablation per pulse, is able to ablate complex (higher order) corneal shapes accurately.

As demonstrated by Pedder et al. [25] and Jiménez et al. [26], the incorporation of models taking into account the angular dependence of laser-ablation rates as well as the effect of plume absorption can be important in efforts to improve the ablation algorithms used in refractive surgery. Differences in corneal power and corneal asphericity encountered when using this model can significantly affect the visual function of patients after LASIK. The high accuracy of determination of stroma plume absorption coefficients and the incorporation of this information in laser-ablation equations can improve the prediction of postsurgical corneal shape. More accurate values for postsurgical radius and asphericity could be achieved and thereby enhance emmetropization and correction of eye aberrations in refractive surgery. However, in case of AMARIS, the effect of the ablation plume could be not so siginificant since a debris removal system is incorporated.

In our study, the range 188–565 mJ/cm2 resulted as optimum fluence level for first simulation, and optimum proportion range 50%–90% high fluence for second one. With optimal parameters, it corresponds to 2.4 s/D at 6 mm OZ, with fidelity variance of 53 m RMS, and average ablation error of 0.5 m for each location.

Ablation simulation of coma Z[], showed 98,4% accuracy, and 98% fit quality; trefoil Z[], 99,9% accuracy and 98% fit quality; spherical aberration Z[], 96,6% accuracy and 97% fit quality; secondary astigmatism Z[], 97,9% accuracy and 98% fit quality. Ablation on a flat plate of PMMA of coma Z[], showed 96,7% accuracy and 96% fit quality; trefoil Z[], 97,1% accuracy and 96% fit quality; spherical aberration Z[], with 93,9% accuracy and 90% fit quality; and secondary astigmatism Z[], with 96,0% accuracy and 96% fit quality.

The proposed dual fluence concept uses a high-fluence “HF” and a low-fluence “LF” level. HF to speed up the treatment (minimised ablation time), LF to ensure the highest accuracy (maximised ablation smoothness). In our simulations results, HF takes out 220 pl vol per pulse (725 nm peak depth per pulse) whereas LF removes 110 pl vol per pulse (450 nm peak depth per pulse).

The amount of treatment that will receive HF pulses is optimised, to keep the overall quality of the ablation as good as possible (dynamically adjusted to speed up the treatment maintaining the highest accuracy). That means that 1 or 2 D will be made only by LF, but a higher diopter treatment will be made up to 95% with HF. Typically, about 80% of the ablation will be performed at HF.

The system uses the “automatic fluence level adjustment” procedure for optimal ablation control. Depending on the planned refractive correction, about 80 percent of the corneal ablation is performed with a high fluence level speeding up the treatment. Fine correction is performed with a low fluence level improving the resolution. The advantage is that the laser treatment is significantly shortened, especially when higher refractive corrections are involved—without compromising on precision and safety.

The analysis of the clinical results specifically addressing customized treatments will show whether there are corneal discrepancies between real and expected shapes. Clinical outcomes published up to now show consistent results [2730].

Ablation of aspherical and customised shapes based upon Zernike polynomials up the the 8th order seems accurate using the dual fluence concept implemented at the AMARIS platform.

In summary, this study demonstrated that it is possible to develop new algorithms and ablation strategies for efficiently performing laser corneal refractive surgery in a customised form. The availability of such profiles, potentially maximising visual performance without incresing the factors of risk, would be of great value for the refractive surgery community and ultimately for the health and safety of the patients.

Further clinical evaluations on human eyes are needed to confirm the preliminary simulated results presented herein.