Journals
Publish with us
Publishing partnerships
About us
Blog
Advances in Pharmaceutics
Table of Contents
Advances in Pharmaceutics
/
2014
/
Article
/
Tab 3
/
Research Article
Development and Validation of Acyclovir HPLC External Standard Method in Human Plasma: Application to Pharmacokinetic Studies
Table 3
Interday precision and accuracy.
Nominal concentration ng/mL
25.0
100
150
Mean
24.92
98.74
149.91
SD
0.02
0.04
0.01
RSD (%)
9.15
7.82
6.10
Nominal (%)
98.93
98.21
99.16
5
5
5