Research Article

Development and Validation of Acyclovir HPLC External Standard Method in Human Plasma: Application to Pharmacokinetic Studies

Table 6

System suitability studies.

Serial numberParametersAcyclovir

1Number of plates/meter4867
2Asymmetric factor0.91
3LOD (ng/mL)8
4LOQ (ng/mL)24