A Rapid, Isocratic HPLC Method for Determination of Insulin and Its Degradation Product

Table 2

Data of precision experiment ( , each preparation was injected twice).

Conc.0.06810 mg/mL0.01703 mg/mL
Ins. areaIS areaARAve.Conc.Ins. areaIS areaARAve.Conc.

9.98688.17341.22191.22600.070702.69688.89420.30320.30570.01786
9.97928.11261.23012.70078.76430.3082
11.79239.79091.20441.20040.069232.74099.09750.30130.30250.01767
11.60489.70041.19632.73489.00550.3037
11.64839.55811.21871.22120.070432.86289.78530.29260.29260.01711
11.68959.55311.22362.86489.78910.2926
10.92149.09791.20041.21170.069883.106610.12970.30670.30810.01800
11.00288.99651.22303.164210.22120.3096
11.23859.25031.21491.22060.070393.195310.66440.29960.29890.01747
11.32899.23811.22633.176110.65100.2982
10.83158.94311.21161.20030.069232.92539.67080.30250.30480.01780
10.63408.94081.18942.97929.70130.3071

Average0.069980.01765
SD0.000640.00032
RSD %0.911.82