Analytical Performances of Human Immunodeficiency Virus Type 1 RNA-Based Amplix® Real-Time PCR Platform for HIV-1 RNA Quantification

Table 1

Lower limit of detection of the HIV-1 RNA-based Amplix real-time PCR assay using the 3rd WHO HIV-1 subtype B international standard (National Institute for Biological Standards and Control) at 185,000 IU/mL (e.g., 5.04 log copies/mL), estimated at 28.1 HIV-1 RNA copies/mL (95% CI, 20.1–82.8 copies/mL)µ.

LevelHIV-1 RNA (copies/mL)Number testedNumber positivityRate (%)

11004141100
2753737100
3504545100
440504896
530656397
620585696
710492755

lower limit of detection was calculated by PROBIT regression analysis using SPSS software.