Comparison of the analytical performance of the different modified electrodes for the determination of DCF.
| Electrodes | Technique | Linear range (μM) | LOD (μM) | References |
| Gold nanoparticle/multiwalled carbon nanotube modified glassy carbon electrode | SWV | 0.03–200 | 0.02 | [9] | Cu-doped zeolite-expanded graphite-epoxy electrode | DPV | 0.3–20 | 0.05 | [36] | APTES-amino-AT-silica/GCE | SWV | 0.3–20 | 0.053 | [37] | Gold nanoparticles decorated multiwalled carbon nanotubes/graphene oxide/AuE | DPV | 0.4–80 | 0.09 | [38] | Ionic liquid modified carbon nanotube paste electrode | DPV | 0.5–300 | 0.2 | [7] | Amino-AT/GCE | SWV | 0.3–20 | 0.204 | [37] | Au-Pt bimetallic nanoparticles decorated multiwalled carbon nanotubes/gold electrode | DPV | 0.5–1000 | 0.3 | [39] | ZIF-67/g-C3N4 | DPV | 0.2–2.2 | 0.071 | The present work |
|
|
Square wave voltammetry: SWV.
|