Research Article

Internally Controlled, Generic Real-Time PCR for Quantification and Multiplex Real-Time PCR with Serotype-Specific Probes for Serotyping of Dengue Virus Infections

Table 3

Dynamic range of generic real-time PCRa.

DENV in vitro RNA copies in extractionDENV in vitro RNA copies in RTDENV in vitro RNA copies in PCRbIC copies in PCRb Number positives DENV in vitro RNA (%; mean Cq-value: SD) Number positives IC (%; mean Cq-value: S.D)

10710650012/12 (100%; 26.59; 0.54)12/12 (100%; 35.82; 1.08)
10610550012/12 (100%; 30.59; 0.46)12/12 (100%; 32.67; 0.79)
10510450012/12 (100%; 34.35; 1.09)12/12 (100%; 32.97; 0.74)
1041035008/12 (66.7%; 37.22; 0.73)12/12 (100%; 33.73; 0.63)
1031025001/12 (8%; 40.71)12/12 (100%; 32.24; 0.60)
00000/120/12

aThe dynamic range of the generic PCR was determined on 10-fold dilution series of DENV in vitro RNA in the presence of a fixed amount of IC spiked into lysis buffer containing DENV negative human serum as background and for 12 replicates per dilution.
bNumber of copies cDNA/PCR has been calculated as 1/12.5 of the extracted RNA. assuming 100% efficiency in extraction, reverse transcription (RT), and PCR.
SD: standard deviation.