Research Article

Simultaneous Identification and Quantification of Canrenone and 11-α-Hydroxy-Canrenone by LC-MS and HPLC-UVD

Table 2

Accuracy of HPLC-UV method for canrenone.

SamplesOriginal (g)Added (g)Found (g)Recoverya (%)Mean Recovery (%)RSDb (%)

10.00140.00970.010795.88
20.00150.00970.010491.75
30.00140.00970.010694.8594.931.21
40.00130.00970.010897.94
50.00140.00970.010593.81

aRecovery (%) = ((canrenone amount found−original canrenone amount in sample solution)/canrenone added amount) × 100%.