Quantification of Cigarette Smoke Particle Deposition In Vitro Using a Triplicate Quartz Crystal Microbalance Exposure Chamber

Table 1

Mean deposited mass values obtained from the single unit QCM [15] ( ) compared to the 3-in-1 QCM unit ( ); the percentage difference between mean mass detected by the two tools; the values indicating no significant difference between tools at all dilutions tested. For the single unit, the values were the mean of 5 repeat experiments with the QCM (in position 2 of the chamber) [15]; for the triplicate QCM the values were the mean of 5 repeat experiments and 3 QCM positions per repeat ( ).

Mean deposited mass ( g/cm2) ± SD
Dilution (1 :  )51025100400

Single unit25.75 ± 2.3010.51 ± 0.423.29 ± 0.240.69 ± 0.090.22 ± 0.03
Triplicate unit28.00 ± 2.2511.51 ± 1.813.64 ± 0.720.75 ± 0.200.21 ± 0.06
Difference (%)8.44.55.24.92.3
value0.0710.2450.3040.5030.573