Clinical Study

Comparison of Neurologic and Radiographic Outcomes with Solitaire versus Merci/Penumbra Systems for Acute Stroke Intervention

Table 6

The outcomes of the present study compared to the results from large mechanical thrombectomy trials.

Present studyMerci [9]Multi-Merci [8]Penumbra pivotal stroke [3]SWIFT trial [10] (solitaire arm)TREVO2 trial [7] (TREVO2 arm)

Sample, 311411641255888
Recanalization, (%)29 (93.5%)*68 (48.2%)112 (68.2%)102 (81.6%)48 (88.9%)76 (86%)
symptomatic ICH, (%)0 (0)11 (7.8%)16 (9.7%)14 (11.2%)1 (1.7%)4 (4%)
3 month mortality, (%)3 (10.3%)61 (43.2%)56 (34.1%)31 (32.8%)10 (17.2%)29 (33%)
3 month mRS ≤ 2, (%)20 (69%)39 (27.6%)59 (35.9% )52 (41.6%)32 (58.2%)38/85 (40%)

This is the final recanalization rate which includes rescue treatment with other devices as well as iaTPA. With Solitaire FR device alone, the recanalization rate was 79.5%.