Development of a Stair-Step Multifrequency Synchronized Excitation Signal for Fast Bioimpedance Spectroscopy
Table 4
Results of impedance measurement experiment under the excitation of the MFS signal
(4,
).
| Harmonics | Frequency (kHz) | Standard | Measured | Relative errors (%) | Standard | Measured | Absolute errors (°) |
| | 16 | 374.39 | 374.32 | 0.02 | −11.48 | −10.91 | 0.57 | | 32 | 326.91 | 326.23 | 0.21 | −17.53 | −17.54 | 0.01 | | 64 | 263.19 | 261.58 | 0.61 | −18.73 | −18.58 | 0.15 | | 128 | 222.37 | 222.16 | 0.09 | −13.50 | −13.56 | 0.06 |
| Mean error | | | | 0.23 | | | 0.20 |
|
|