Research Article

The Construct Validity and Reliability of an Assessment Tool for Competency in Cochlear Implant Surgery

Table 2

Interrater agreement for task-based checklist.

ItemsIntraclass correlation coefficient (ICC)95% CI value

Left slightly overhanging superiorly, posteriorly, and inferiorly to help retain and control the electrode cable0.860.66 to 0.94<0.001
Thin posterior EAC cortex maximally0.690.31 to 0.860.003
Preserve a layer of bone overlying the facial nerve0.850.68 to 0.93<0.001
Remove bone on the anterior medial surface of the fallopian canal just below the pyramidal process0.740.44 to 0.88<0.001
Preserve a layer of bone overlying the chorda tympani0.840.66 to 0.93<0.001
Facial recess is opened to visualize the round window niche0.930.84 to 0.97<0.001
Cochleostomy is drilled through the promontory just anterior and inferior to the round window membrane0.790.53 to 0.910.037
The bony promontory overlying the round window niche has been removed

All variables0.920.86 to 0.96<0.001