Definition of Metrics to Evaluate Cochlear Array Insertion Forces Performed with Forceps, Insertion Tool, or Motorized Tool in Temporal Bone Specimens

Table 1

Metric values recorded during of the cochlear implantation with three different insertion techniques.

MetricInsertion techniqueMean ± SD

Peak
(N)
Forceps0.256 ± 0.06110NA
Hifocus tool0.327 ± 0.05580.028
Motorized0.255 ± 0.0758NS

Momentum
(Ns)
Forceps1.16 ± 0.50510NA
Hifocus tool1.337 ± 0.4088NS
Motorized 1.573 ± 0.7648NS

Over threshold
>0.1 N
Forceps21.00 ± 12.55210NA
Hifocus tool7.00 ± 4.03680.002
Motorized 3.38 ± 3.11380.0002

Sudden riseForceps90.60 ± 46.56910NA
Hifocus tool28.25 ± 15.87280.0003
Motorized 14.00 ± 6.94980.00003

Jerk
(N·s−1)
Forceps0.467 ± 0.11610NA
Hifocus tool0.515 ± 0.2068NS
Motorized 0.1553 ± 0.0580.00008

Values are expressed as mean ± SD of insertion. NA: not applicable; NS: not significant. “Forceps” stands for manual insertion with forceps technique, “Hifocus tool” stands for Hifocus 1J electrode insertion tool technique, and “motorized” stands for our in-house motorized insertion tool technique. Statistical analysis was performed by analysis of variance. Each technique was compared against the manual insertion with forceps technique.