Table of Contents Author Guidelines Submit a Manuscript
BioMed Research International
Volume 2014, Article ID 586060, 5 pages
http://dx.doi.org/10.1155/2014/586060
Research Article

Effect of Intraocular Lens Diameter Implanted in Enucleated Porcine Eye on Intraocular Pressure Induced by Scleral Depression

1Department of Ophthalmology, Teikyo University School of Medicine, 2-11-1 Kaga, Itabashi-ku, Tokyo 173-8605, Japan
2Matsumoto Eye Clinic, 50-2 Takagaki, Awa-cho, Awa-shi, Tokushima 771-1705, Japan

Received 28 December 2013; Accepted 2 March 2014; Published 27 March 2014

Academic Editor: Toshihide Kurihara

Copyright © 2014 Gaku Terauchi et al. This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

Abstract

The effect of the diameter of an intraocular lens (IOL) implanted in enucleated porcine eyes on the intraocular pressure induced by scleral depression was investigated. Two IOLs of 6 mm and 7 mm optic diameter were implanted. The intraocular pressure (IOP) was monitored during scleral depression by a transducer placed in the midvitreous through a sclerotomy at 6 o’clock. The area under the curve (AUC) of the IOP changes from the beginning of the indentation to the point when the peripheral retinal surface was observed through the IOL optics was measured. The AUC was significantly larger in eyes with a 6 mm IOL than in eyes with a 7 mm IOL (). The IOP elevation at the endpoint was higher in eyes with the 6 mm IOL than in eyes with the 7 mm IOL. We conclude that the AUC may represent the degree of stress induced by scleral depression. The higher AUC value with the X-60 may be because of the longer distance from the peripheral retina to the edge of the IOL optics.