BioMed Research International / 2014 / Article / Tab 1 / Research Article
Serial Changes of Neointimal Tissue after Everolimus-Eluting Stent Implantation in Porcine Coronary Artery: An Optical Coherence Tomography Analysis Table 1 Quantitative coronary angiographic and optical coherence tomographic findings of the entire stent segment at 1, 3, and 6 months after stent implantation.
Bare-metal stent ( ) Everolimus-eluting stent ( ) 1 month 3 months 6 months 1 month 3 months 6 months Quantitative coronary angiography analyses Proximal reference, mm 2.7 (2.6–3.1) 2.8 (2.6–3.2) 2.8 (2.5–3.2) 2.9 (2.7–3.2) 3.0 (2.6–3.1) 2.7 (2.4–3.1) Distal reference, mm 2.6 (2.4–2.8) 2.6 (2.3–3.0) 2.5 (2.2–2.9) 2.6 (2.4–2.8) 2.5 (2.4–2.7) 2.5 (2.3–2.7) Minimal lumen diameter, mm 2.2 (2.0–2.5) 2.2 (2.1–2.4) 2.2 (2.0–2.4) 2.4 (2.0–2.6)* 1.7 (1.4–2.3) 1.8 (1.5–2.2) Optical coherence tomography findings Neointimal thickness, m 190.4 (127.0–243.9) 182.2 (131.9–229.4) 169.8 (112.7–211.1) 120.6 (93.8–203.9)* 276.5 (172.2–428.1) 269.6 (199.9–396.4) Stent area, mm2 6.1 (5.2–6.8) 6.1 (5.4–7.0) 6.0 (5.2–6.8) 6.2 (5.3–6.9) 6.3 (5.4–7.0) 6.3 (5.0–6.9) Lumen area, mm2 4.5 (3.8–5.6) 4.7 (4.0–5.4) 4.6 (3.8–5.3) 4.9 (4.1–5.8)* 3.8 (3.0–4.9) 3.8 (3.2–4.5) Neointimal area, mm2 1.6 (1.1–1.9) 1.4 (1.0–1.9) 1.4 (0.9–1.7) 1.3 (0.9–1.7)* 2.5 (1.4–3.6) 2.4 (1.6–3.1) Percentage of neointimal hyperplasia area, % 25.3 (19.2–29.4) 23.9 (17.3–32.0) 23.2 (13.1–29.9) 21.5 (12.9–28.1)* 39.1 (24.1–53.7) 35.4 (31.7–45.0)
, using Friedman tests.