Serial Changes of Neointimal Tissue after Everolimus-Eluting Stent Implantation in Porcine Coronary Artery: An Optical Coherence Tomography Analysis

Table 1

Quantitative coronary angiographic and optical coherence tomographic findings of the entire stent segment at 1, 3, and 6 months after stent implantation.

Bare-metal stent ()Everolimus-eluting stent ()
1 month3 months6 months1 month3 months6 months

Quantitative coronary angiography analyses
 Proximal reference, mm2.7 (2.6–3.1)2.8 (2.6–3.2)2.8 (2.5–3.2)2.9 (2.7–3.2)3.0 (2.6–3.1)2.7 (2.4–3.1)
 Distal reference, mm2.6 (2.4–2.8)2.6 (2.3–3.0)2.5 (2.2–2.9)2.6 (2.4–2.8)2.5 (2.4–2.7)2.5 (2.3–2.7)
 Minimal lumen diameter, mm2.2 (2.0–2.5)2.2 (2.1–2.4)2.2 (2.0–2.4)2.4 (2.0–2.6)*1.7 (1.4–2.3)1.8 (1.5–2.2)
Optical coherence tomography findings
 Neointimal thickness, m190.4 (127.0–243.9)182.2 (131.9–229.4)169.8 (112.7–211.1)120.6 (93.8–203.9)*276.5 (172.2–428.1)269.6 (199.9–396.4)
 Stent area, mm26.1 (5.2–6.8)6.1 (5.4–7.0)6.0 (5.2–6.8)6.2 (5.3–6.9)6.3 (5.4–7.0)6.3 (5.0–6.9)
 Lumen area, mm24.5 (3.8–5.6)4.7 (4.0–5.4)4.6 (3.8–5.3)4.9 (4.1–5.8)*3.8 (3.0–4.9)3.8 (3.2–4.5)
 Neointimal area, mm21.6 (1.1–1.9)1.4 (1.0–1.9)1.4 (0.9–1.7)1.3 (0.9–1.7)*2.5 (1.4–3.6)2.4 (1.6–3.1)
 Percentage of neointimal hyperplasia area, %25.3 (19.2–29.4)23.9 (17.3–32.0)23.2 (13.1–29.9)21.5 (12.9–28.1)*39.1 (24.1–53.7)35.4 (31.7–45.0)

, using Friedman tests.