Research Article

Dose-Response-Relationship between Number of Laser Burns and IOP Reduction in Cyclophotocoagulation: An Animal Study

Table 1

IOP differences between treated and control eye ( IOP in mmHg), before and after treatment, and calculated IOP reduction per laser burn. A negative difference means that the IOP of the treated eye is lower than that of the control eye.

Laser burns15102030
Before treatment0.1 ± 0.20 ± 0.3−0.2 ± 0.3−0.3 ± 0.2−0.3 ± 0.2
After treatment−1.0 ± 0.4−2.6 ± 0.6−3.4 ± 0.8−4.7 ± 1.0−6.1 ± 1.4
value0.02<0.010.010.010.01
IOP reduction per laser burn0.96 ± 0.370.52 ± 0.130.34 ± 0.080.24 ± 0.050.20 ± 0.05