Mean Expected Error in Prediction of Total Body Water: A True Accuracy Comparison between Bioimpedance Spectroscopy and Single Frequency Regression Equations

Table 5

Bland-Altman analysis for comparison of total body water predicted by BIS and 50 kHz impedance methods.

Prediction methodVolume (L) Bias1CorrelationSEE
Liters
(mean ± SD)
%
(mean ± SD)
PearsonLin

Reference method 2H2O dilution40.47 ± 9.2

Deurenberg et al. [18]38.6 ± 7.72.3 ± 4.14.63 ± 9.70.900.854.65
Heitmann [19]41.35 ± 6.9−0.88 ± 4.2−3.97 ± 11.10.900.864.25
Kushner and Schoeller [20]43.17 ± 9.0−2.70 ± 2.4−7.35 ± 6.90.970.923.62
Sun et al. [21]42.56 ± 9.7−2.09 ± 2.6−5.37 ± 6.30.970.943.28

BIS prediction methods
Moissl et al. [24]39.9 ± 8.50.60 ± 2.4−0.99 ± 5.70.970.962.48
de Lorenzo et al. [22]39.6 ± 9.220.86 ± 2.5−2.03 ± 5.90.960.962.59

Compared to reference method.