Research Article

Comparative Study of Anterior Eye Segment Measurements with Spectral Swept-Source and Time-Domain Optical Coherence Tomography in Eyes with Corneal Dystrophies

Table 2

Bland-Altman plot comparing automated and manual central corneal thickness (aCCT, mCCT), anterior chamber depth (ACD), and nasal and temporal trabecular iris angle (nTIA, tTIA) measurements by swept-source optical coherence tomography SS OCT and time-domain optical coherence tomography TD OCT in control and corneal dystrophies group (CD group). 95% CI:95% confidence interval of the mean difference; SD: standard deviation.

Control groupCD group
Mean difference95% CISDMean difference95% CISD

aCCT−4.4−10.47–1.6321.29.14−4.32−8.51–0.1320.35.04
mCCT−3.4−7.7–0.9715.37.12−3.75−8.81–1.3124.58.14
ACD0.070.06–0.090.06<.0010.06−0.01–0.140.38.01
nTIA−1.58−2.48–−0.673.2.0011.80.14–3.476.62.03
tTIA−2.22−3.04–−0.392.91<.001−1.97−2.59–−1.362.97<.001