A Novel Simple Phantom for Verifying the Dose of Radiation Therapy
Table 1
The dose comparison between ionization chamber and TLDs.
| | (cGy) | (cGy) | Difference (%) |
| 6 MV photon | 194.48 ± 2.32 | 199.50 ± 1.92 | −2.519% | 10 MV photon | 197.21 ± 3.93 | 197.97 ± 1.87 | −0.386% | 6 MeV electron | 203.71 ± 2.04 | 199.32 ± 1.56 | 2.205% | 9 MeV electron | 194.84 ± 3.62 | 198.45 ± 2.03 | −1.819% | 12 MeV electron | 198.99 ± 3.13 | 198.04 ± 1.58 | 0.478% |
|
|
: dose measured from TLDs; : dose measured from the ionization chamber.
|