A Novel Simple Phantom for Verifying the Dose of Radiation Therapy

Table 1

The dose comparison between ionization chamber and TLDs.

(cGy) (cGy)Difference (%)

6 MV photon194.48 ± 2.32199.50 ± 1.92−2.519%
10 MV photon197.21 ± 3.93197.97 ± 1.87−0.386%
6 MeV electron203.71 ± 2.04199.32 ± 1.562.205%
9 MeV electron194.84 ± 3.62 198.45 ± 2.03−1.819%
12 MeV electron198.99 ± 3.13198.04 ± 1.580.478%

: dose measured from TLDs; : dose measured from the ionization chamber.