Research Article

A Randomized Comparison Simulating Face to Face Endotracheal Intubation of Pentax Airway Scope, C-MAC Video Laryngoscope, Glidescope Video Laryngoscope, and Macintosh Laryngoscope

Table 4

Success rate according to the number of attempts.

ā€‰MCLAWSGVLCMAC

Success at 1st attempt72 (83.7%)71 (82.5%)37 (43.0%)74 (86.0%)
Success at 2nd attempt13 (15.1%)
(85, 98.8%)
13 (15.1%)
(84, 97.6%)
25 (29.0%)
(62, 72.0%)
9 (10.4%)
(83, 96.5%)
Success at 3rd attempt1 (1.1%)
(86, 100%)
1 (1.1%)
(85, 98.8%)
8 (9.3%)
(70, 81.3%)
2 (2.3%)
(85, 98.8%)
Success at 4th attempt01 (1.1%)
(86, 100%)
3 (3.4%)
(73, 84.8%)
0

Failure at 4th attempt0013 (15.1%)1 (1.1%)

MCL: Macintosh laryngoscope; AWS: Pentax airway scope; GVL: Glidescope video laryngoscope; C-MAC: C-MAC video laryngoscope.