Accuracy and Precision of Three-Dimensional Low Dose CT Compared to Standard RSA in Acetabular Cups: An Experimental Study

Table 1

Precision of 3D CT compared to standard RSA.

Standard RSA3D CT
2.45 SDMeanMinMax2.45 SDMeanMinMax

Uncemented cup
 Translation (mm)
  0.070.00−0.030.050.010.000.000.01
  0.050.02−0.020.040.04−0.01−0.040.00
  0.090.03−0.020.090.09−0.01−0.090.01
 Rotation (°)
  0.22−0.05−0.150.120.10−0.01−0.100.01
  0.130.00−0.100.050.06−0.02−0.050.00
  0.080.00−0.030.050.21−0.07−0.210.01
Cemented cup
 Translation (mm)
  0.080.00−0.040.050.04−0.01−0.030.00
  0.040.00−0.020.020.04−0.01−0.040.00
  0.090.00−0.050.050.06−0.02−0.060.00
 Rotation (°)
  0.19−0.03−0.160.060.14−0.02−0.110.02
  0.320.02−0.170.140.29−0.14−0.240.02
  0.120.02−0.030.100.27−0.05−0.270.02