Research Article

In Vivo Corneal Biomechanical Properties with Corneal Visualization Scheimpflug Technology in Chinese Population

Table 1

Repeatability of the recalculated corneal visualization Scheimpflug technology parameters for three measurements.

ParametersPrecisionRepeatabilityCV%ICC95% CI for ICC

First applanation time (msec)0.1870.2641.3060.9290.901~0.950
First applanation length (mm)0.1130.1603.2210.2230.094~0.362
Velocity in (m/sec)0.0180.0266.3530.6950.601~0.776
Second applanation time (msec)0.2840.4010.6570.9290.901~0.951
Second applanation length (mm)0.5400.76416.1530.2300.101~0.369
Velocity out (m/sec)0.0760.107−9.8340.7680.690~0.832
Highest concavity time (msec)0.6840.9662.0930.3020.171~0.438
Deformation amplitude (mm)0.0690.0983.3120.9420.920~0.960
Peak distance (mm)0.1680.2381.7010.8960.857~0.927
Highest concavity radius (mm)1.1361.6068.3200.6050.495~0.704
Intraocular pressure (mmHg)1.6962.3976.2450.8940.854~0.925
Central corneal thickness (μm)9.64213.6270.9250.9830.976~0.988

. CV: coefficient of variation; ICC: intraclass correlation coefficient; 95% CI: 95% confidence interval of the mean.