TY - JOUR A2 - Michalakis, Konstantinos AU - Ogasawara, Takeshi AU - Uezono, Masayoshi AU - Takakuda, Kazuo AU - Kikuchi, Masanori AU - Suzuki, Shoichi AU - Moriyama, Keiji PY - 2017 DA - 2017/12/17 TI - Shape Optimization of Bone-Bonding Subperiosteal Devices with Finite Element Analysis SP - 3609062 VL - 2017 AB - Subperiosteal bone-bonding devices have been proposed for less invasive treatments in orthodontics. The device is osseointegrated onto a bone surface without fixation screws and is expected to rapidly attain a bone-bonding strength that successfully meets clinical performance. Hence, the device’s optimum shape for rapid and strong bone bonding was examined in this study by finite element analyses. First, a stress analysis was performed for a circular rod device with an orthodontic force parallel to the bone surface, and the estimate of the bone-bonding strength based on the bone fracture criterion was verified with the results of an animal experiment. In total, four cross-sectional rod geometries were investigated: circular (Cr), elliptical (El), semicircular (Sc), and rectangular (Rc). By changing the height of the newly formed bone to mimic the progression of new bone formation, the estimation of the bone-bonding strength was repeated for each geometry. The rod with the Rc cross section exhibited the best performance, followed by those with the Sc, El, and Cr cross sections, from the aspects of the rapid acquisition of strength and the strength itself. Thus, the rectangular cross section is the best for rod-like subperiosteal devices for rapid bone bonding. SN - 2314-6133 UR - https://doi.org/10.1155/2017/3609062 DO - 10.1155/2017/3609062 JF - BioMed Research International PB - Hindawi KW - ER -