Research Article

Determination of Aflatoxin M1 and Chloramphenicol in Milk Based on Background Fluorescence Quenching Immunochromatographic Assay

Table 3

Comparison of test efficiency between LC-MS and bFQICA.

SamplebFQICA (μg/kg)LC-MS (μg/kg)

AFM1 in sample 010.11510.0988
AFM1 in sample 020.14920.151
CAP in sample 030.17680.160
CAP in sample 040.49250.500

bFQICA, background fluorescence quenching immunochromatographic assay; LC-MS, liquid chromatography-mass spectrometry. Two samples were used for the AFM1 (samples 01 and 02) and CAP (samples 03 and 04), respectively. No statistical difference was noticed between the efficiency of bFQICA and LC-MS.