Results coming from the validation of the Raman method developed to quantify taxane drugs in sodium chloride solution.

CBXPCXDCX

Range of linearity (mg/mL)0.175 – 0.300.57 – 1.200.40 – 1.00
Slope1.00010.95980.9917
Intercept7.10−60.02520.0048
R20.99990.95980.9909
LOQ (mg/mL)0.160.560.40

Relative bias (%)
Level 10.36.1−1.4
Level 2−2.25.1−4.2
Level 3−0.2−1.2−5.3
Repeatability (RSD %)
Level 17.62.63.2
Level 22.51.72.3
Level 35.41.71.6
Intermediate precision (RSD %)
Level 110.78.65.2
Level 23.91.82.7
Level 35.41.82.5
β-Expectation tolerance limit (%)
Level 1[−24.7;25.3][−22.8; 35.0][14.5; 11.7]
Level 2[−11.6; 7.2][1.5; 8.8][−9.8; 1.5]
Level 3[−10.8; 10.4][−4.7; 2.2][−11.1; 0.4]