| Specific Indications |
Part of the scanner | TRIOS® | iTero® |
|
System or Base Unit [20] | (i) Surface disinfection | (i) Surface disinfection. |
|
Monitor [20] | (i) Do not spray directly with disinfectant. | (i) Do not spray directly with disinfectant. (ii) Use disinfectant wipes for the Scanning Unit and Base Unit. |
|
Handheld scanner [20] | (i) Do not submerge the handheld scanner in any liquids. (ii) Do not place the handheld scanner on heated or wet surfaces. (iii) Surface disinfection. | Not indicated in open source [21]. |
|
Medical-grade peripherals (e.g., keyboards and mice) [20] | (i) Easy disinfection. | Not indicated in open source [21]. |
|
Scanner tips with fixed mirror or detachable mirror [19] | | Not needed. |
| Immediately after clinical use: (i) Detach the mirror from detachable mirror and go on reconditioning separately for tip and mirror. (ii) Go on reconditioning for tip with fixed mirror. | |
| (iii) Clean manually and perfectly using soapy water and a soft dish brush. (iv) Rinse carefully the tip. (v) Inspect the mirror of the tip after cleaning. (vi) Dry the mirror carefully with a paper towel. (vii) Check to make sure it is free of lint, stains, and other kinds of dirt. | |
| (a) Wrap the tip using a self-adhesive pouch or heat-sealed pouch. | |
| (b) Sterilization using a steam autoclave class B (EN13060) and cycles at 121/134°C with drying. | |
| (c) Storage in proper condition. | |
|
Disposable plastic sleeve [21] | | (i) Dispose of scanner sleeves according to standard operating procedures or local regulations for the disposal of contaminated medical waste. |
|
Type of disinfectant [19] | (i) For optical windows and scanner tips: denatured alcohol (ethyl alcohol or ethanol) – typically 60-70% Alc/Vol. (ii) Mixture free of impurities that can stain the mirror. | (i) Many commercial products. (ii) Follow the disinfectant manufacturers’ instructions for appropriate contact time. (iii) Remove residual liquid disinfectant with a lint-free, clean cloth. |
|
Wipe [19, 20] | A soft lint-free nonabrasive cloth. | (i) Disinfectant wipes. |
|
Prohibition on mirror tip of the use of [19] | (i) Ammonia-based or chloride based solutions or acetone on any surface. (ii) Acetone or any oxidizing solutions to clean the tip mirror. | |
|
Disposal of scanner tip | Normally as other clinical waste. | Normally as other clinical waste. |
|