Hai Yen Mai, Du-Hyeong Lee, "Impact of Matching Point Selections on Image Registration Accuracy between Optical Scan and Computed Tomography", BioMed Research International, vol. 2020, Article ID 3285431, 7 pages, 2020. https://doi.org/10.1155/2020/3285431

Impact of Matching Point Selections on Image Registration Accuracy between Optical Scan and Computed Tomography

Guest Editor: Henriette Lerner
Received17 Jun 2020
Accepted21 Jul 2020
Published05 Aug 2020

Abstract

The point-based surface registration method involves the manual selection process of paired matching points on the data of computed tomography and optical scan. The purpose of this study was to investigate the impact of selection error and distribution of fiducial points on the accuracy of image matching between 3-dimensional (3D) images in dental planning software programs. Computed tomography and optical scan images of a partial edentulous dental arch were obtained. Image registration of the optical scan image to computed tomography was performed using the point-based surface registration method in planning software programs under different conditions of 3 fiducial points: point selection error (0, 1, or 2 mm), point distribution (unilateral, bilateral), and planning software (Implant Studio, Blue Bio Plan) ( per condition, ). The accuracy of image registration at each condition was evaluated by measuring linear discrepancies between matched images at , , and axes. Kruskal-Wallis test, Mann-Whitney test with Bonferroni correction, and 3-way analysis of variance were used to statistically analyse the measurement data (). No statistically significant difference was exhibited between the 0 and 1 mm point mismatch conditions in either unilateral or bilateral point distributions. The discrepancy values in the 2 mm mismatch condition were significantly different from the other mismatch conditions, especially in the unilateral point distribution (). Strong interactions among point selection error, distribution, and software programs on the image registration were found (). Minor matching point selection error did not influence the accuracy of point-based automatic image registration in the software programs. When the fiducial points are distributed unilaterally with large point selection error, the image matching accuracy could be decreased.

1. Introduction

Three-dimensional (3D) imaging technologies have enhanced the diagnostic modalities and treatment planning for implant, maxillofacial surgery, and orthodontic fields [1, 2]. Cone-beam computed tomography (CBCT) is representative, and the 3D image data that it produces are commonly used to evaluate the underlying bone and identifying critical anatomical structures, such as the inferior alveolar nerve, the maxillary sinus, and the roots of neighbouring teeth [3]. Slices of CBCT radiographic images can be reconstructed into a 3D image model [4], but the resolution of the 3D image is limited because of the voxel size of raw radiographic data available in CBCT devices [5, 6]. Thus, to make a 3D model with soft and hard tissue, image merging with optical scan data of the oral cavity surface is recommended [7, 8].

Image registration is the process of matching the optical scan image to the 3D-reconstructed CBCT image [9, 10]. Accurate image registration is essential to replicate the exact relationship of underlying bone and oral cavity surface data [11, 12]. Errors in the image alignment in this step could cause unexpected surgical complications because of discrepancies between the planning in the virtual model and the actual results at the surgical site [13]. Contemporary image registration techniques for 3D data are divided into voxel-based and surface-based method [14]. The voxel based-registration utilizes the grey-scale difference of the voxels to align the images, achieving the least difference in the total grey-scale density [14]. Because the voxel-based registration uses the image intensity in the entire volume, the presence of image artifacts caused by metallic prostheses in the radiographic data could deteriorate precise image matching by masking the fiducial anatomic structures [5]. Meanwhile, the surface-based registration uses the geometric shape of 3D object surfaces to match images [15]. The computer algorithm of iterative closest points (ICPs) locates optimal positions of pair images by computing the 3D coordinates of image surface point clouds [14]. The accuracy of image superimposition by graphic processing has been validated [16], and currently, point-based surface matching is widely used in the dental planning and inspection computer software programs for the enhancement of efficiency and accuracy of matching [9].

When performing the point-based surface registration, the congruent areas between the 3D reconstructed radiographic image and optical scan image are designated in pair [17]. In the dentate jaws, because the tooth is discernible in both images, the outline of the tooth is generally used as the fiducial area of image superimposition [17]. Three pair points have been used for image matching [18], and wide distribution of points in the arch is recommended [19]. It was also suggested to choose the matching points close to the edentulous area [19, 20]. Although several protocols for image registration have been suggested, the paired points could be selected differently because this process basically depends on the operator’s visual observations. Human error in selecting matching points could affect the accuracy of subsequent automatic image matching processes. The recommended positions of the matching points also have not been clarified. The purpose of this study was to assess the impact of selection error and distribution of fiducial points on the accuracy of image matching between 3D radiographic and optical scan images in planning software programs. The proposed null hypothesis was that the mismatch in paired fiducial points, and the different distribution of points would not influence the accuracy of image registration of the 3D optical scan image to the reconstructed radiographic data.

2. Materials and Methods

The workflow of this study is described in Figure 1. A patient missing the second premolar and first molar on the maxilla was selected for this study. The patient had no severe defect or metal restorations in the remaining teeth on the arch. Computed tomographic images of the patient were obtained by using a CBCT scanner (Pax-i3D Smart, Vatech, Hwaseong, Korea) with 80 kVp, 8 mA, 24 s pulsed scan, field of view of  mm, and slice thickness of 0.2 mm. An optical scan image of the oral cavity surface was obtained by digitizing the stone model using a lab-based scanner (IDC S1, Amann Girrbach, Koblach, Austria). The stone model was fabricated using the conventional silicone impression and stone pouring technique. The scan file was transferred to a reverse engineering software program (Geomagic Design X, 3D Systems, Rock Hill, SC, USA) where indexing markers for indicating matching points were made on the occlusal surface of teeth by deleting the surface of the image in the strip shape at 1 mm intervals (Figure 2). Consequently, the radiographic and optical-scan data were delivered to planning computer software programs, Implant Studio (3Shape, Copenhagen, Denmark) and Blue Sky Plan (Blue Sky Bio, LLC, Grayslake, IL, USA), for merging the images.

The experimental factors of this study were selection error in pair points and the distribution of fiducial points in the image registration process. The selection error was set at 3 levels (0, 1, and 2 mm; Figure 3), and the distribution of points was set at 2 levels (unilateral and bilateral; Figure 4). In all matching conditions, 3 matching points were used in pairs in both radiographic and optical-scan images, and the positions of points were designated in reference to the indexing markers to provide the standardized error in point selection [17, 20]. The central incisor, first premolar, and second premolars were used for the image matching in the unilateral and bilateral distribution conditions. After the point designation in each condition, the point-based best-fit algorithm was run in the software programs. All image registrations were carried out 5 times in each condition by a single operator who had experience in image registration and was blinded to the purpose of this study.

The resulting accuracy of each condition’s image registration was evaluated by measuring the positional discrepancy between the radiographic and optical-scan images in the -, -, and -axes (Figure 5). The assessments were performed in the cross-sectional images of central incisor and second molar areas using the measurement function of the planning software programs. A single investigator carried out all the measurements to avoid errors that can arise when using different examiners.

The mean and standard deviation of linear discrepancies in each condition were calculated by averaging the measurement values collected in the anterior and posterior areas. Kruskal-Wallis test and a post hoc Mann-Whitney test with Bonferroni correction were used to compare the accuracy of image registration between groups. Three-way analysis of variance (ANOVA) was used to investigate how interactions between factors, such as the point selection error, point distribution, and planning software, affected image registration accuracy. All statistical analyses were performed by using the Statistical Package for the Social Sciences (SPSS) software program (SPSS version 25.0; IBM Inc., Armonk, NY, USA) with the statistical significance level of 0.05.

3. Results

The linear discrepancy between the radiographic and optical-scan images after the image registration process at different matching conditions is shown in Table 1. In general, no statistically significant difference was found between the 0 and 1 mm point mismatch conditions; however, the discrepancy values in the 2 mm mismatch condition were significantly different from the other mismatch conditions. In particular, when the 2 mm point mismatch condition was applied in the unilateral point distribution in the Blue Sky Plan software, markedly high discrepancy was observed (9.92 mm in the -axis, 2.52 mm in the -axis, and 4.19 mm in the -axis).


CoordinateDistributionImplant studioBlue Sky Plan
Point mismatchPoint mismatch
012012

Bilateralaab3b0.101aabb0.055
Unilateralaab<0.001aab<0.001
0.205.141.0060.096.543<0.001
Bilateral0.2580.901
Unilateralaab<0.001a0ab<0.001
0.9510.651<0.0010.9360.689<0.001
Bilateralabb<0.001abc<0.001
Unilateralaab0.002aab<0.001
0.0670.5760.0360.0900.052<0.001

Values with the same letter are not statistically different based on Kruskal-Wallis test at .

The pattern of point distribution did not make a statistically significant difference in the matching discrepancy in the 0 and 1 mm point mismatch conditions, but it did make significant differences in the 2 mm mismatch condition in both planning software programs.

Three-way ANOVA results showed that each factor has strong interactions with different factors, which affects the outcome of image matching, as shown in Table 2 (; adjusted ).


SourceType III sum of squaresd.f.Mean square

Corrected model1142.82711103.893370.319<0.001
Intercept209.6931209.693747.433<0.001
Distribution123.9361123.936441.758<0.001
Software76.649176.649273.210<0.001
Mismatch257.1512128.576458.297<0.001
241.6872120.844430.737<0.001
174.877287.439311.668<0.001
90.653190.653323.124<0.001
177.873288.937317.007<0.001
Error47.1331680.281
Total1399.652180
Corrected Total1189.959179

Adjusted .

4. Discussion

The purpose of this study was to elucidate the impact of the matching point selection process on the accuracy of optical scan image registration to radiographic images for image merging. Variations in the process of manual point designation were designed by controlling the factors, such as the disagreement in paired points and the distribution of points. The overall results of this study showed that minor errors in matching point selection did not influence the accuracy of image registration regardless of whether the distribution pattern was unilateral or bilateral. However, the discrepancy was significantly different when there was large point mismatch. Thus, the null hypothesis that the mismatch in paired fiducial points and the different distribution of points would not influence the accuracy of image registration was rejected.

The image matching method used in the study was point-based surface registration. Polygonal 3D mesh images were aligned with each other in the closest position using the best-fit algorithm based on the fiducial points designated by the operator [21]. The results of this study showed that minor errors in the point selection did not affect the quality of image superimposition of paired images. This accurate matching algorithm not only reduces the influence of iatrogenic mistakes in the image superimposition for treatment planning but also can be used for result analyses and research purposes in various medical fields [22, 23]. Because the matching method is based on mathematical calculations, it is assumed that there is no matching error when the matching is performed in ideal conditions. However, in the present study, there was some discrepancy when there was no point selection error. This phenomenon might be due to the differences in forms of 3D restructured radiographic and optical scan images. Shape deformation has been reported in the conversion process from radiographic raw data to 3D mesh image because of the partial volume effect [2426]. The optical scan model could also have dimensional errors depending on scanner performance and scanning strategy [27, 28]. Thus, operators should be aware of the errors that inevitably occur because of the morphological differences between matched images, and they should focus on minimizing additional operator errors caused by inappropriate manual works.

In this study, point-based surface registration was performed in planning software programs that are widely used in guided implant surgery [9]. The general results were similar in the 2 software programs. However, the matching results were different in unilateral point distribution with 2 mm selection error. In the Implant Studio, the optical scan image was approximately located near the corresponding radiographic image, but in the Blue Sky Plan, the image matching failed completely. The malfunctioning of the best-fit algorithm might be due to the fact that the level of point mismatch in a certain point distribution was beyond the condition that is needed for the automatic image matching to operate normally. Based on the findings of the present study, it seems that the maximum allowance limit that enables normal operation of best-fit image matching might be different according to different computer software programs. Scherer [23] reported that the accuracy of outcomes could vary depending on the software programs used. Further studies on the capability of image matching in different commercial software programs are needed.

The conventional procedures of silicone impression taking and gypsum cast fabrication in this study could be sources of error in replicating the morphology of the oral cavity because of the physical and chemical characteristics of the materials [29]. Intraoral optical scanning can be considered to directly obtain image data from the oral cavity, eliminating the drawbacks of analog methods [30]. A partially edentulous dental arch case was chosen for this study. Considering the clinical variability in the size of dental arches, positions and spans of the edentulous area, and tooth morphology, various clinical cases should be included in future studies to generalize the results of the present study. Cases with metallic restorations also need to be included to investigate the effects of metal artifact images on the accuracy of image registration. The proficiency of the operator in the image matching could be another influencing factor. The operator factor was controlled in the present study by recruiting operators who had the same amount of experience in using planning computer software. Clinically, given that operator’s experience varies, it will be necessary to examine the relationship between experience and point selection error in related future studies.

5. Conclusions

Within the limitations of this study, the following conclusions were drawn: (1)A matching point selection error of 1 mm did not affect the accuracy of optical scan image registration to radiographic image in either unilateral or bilateral point distributions(2)Whether the fiducial points were distributed unilaterally or bilaterally did not affect the accuracy of image registration when there was no or 1 mm point selection error(3)The accuracy of image registration was significantly different between software programs when the fiducial points were distributed unilaterally with a 2 mm selection error

Data Availability

The data used to support the findings of this study are included within the article.

Conflicts of Interest

The authors declare that there is no conflict of interest regarding the publication of this paper.

Authors’ Contributions

HY Mai contributed to the conceptualization of the study, data curation, formal analysis, investigation, and writing of the manuscript. DH Lee was involved in the methodology, data curation, supervision, and critical revision of the manuscript. All authors have read and approved the final manuscript.

Acknowledgments

The authors thank Hang-Nga Mai and Yong-Do Choi for obtaining data for this experiment.

References

  1. G. Orentlicher, D. Goldsmith, and A. Horowitz, “Applications of 3-dimensional virtual computerized tomography technology in oral and maxillofacial surgery: current therapy,” Journal of Oral and Maxillofacial Surgery, vol. 68, no. 8, pp. 1933–1959, 2010. View at: Publisher Site | Google Scholar
  2. E. Honda, J. L. Prince, and V. R. C. Fontanella, “State-of-the-art digital imaging in dentistry: advanced research of MRI, CT, CBCT, and digital intraoral imaging,” BioMed Research International, vol. 2018, Article ID 9057120, 2018. View at: Publisher Site | Google Scholar
  3. M. Bornstein, W. Scarfe, V. Vaughn, and R. Jacobs, “Cone beam computed tomography in implant dentistry: a systematic review focusing on guidelines, indications, and radiation dose risks,” International Journal of Oral and Maxillofacial Implants, vol. 29, Supplement, pp. 55–77, 2014. View at: Publisher Site | Google Scholar
  4. T. Kulczyk, M. Rychlik, D. Lorkiewicz-Muszyńska, M. Abreu-Głowacka, A. Czajka-Jakubowska, and A. Przystańska, “Computed tomography versus optical scanning: a comparison of different methods of 3D data acquisition for tooth replication,” BioMed Research International, vol. 2019, Article ID 4985121, 7 pages, 2019. View at: Publisher Site | Google Scholar
  5. G. R. J. Swennen, E.-L. Barth, C. Eulzer, and F. Schutyser, “The use of a new 3D splint and double CT scan procedure to obtain an accurate anatomic virtual augmented model of the skull,” International Journal of Oral and Maxillofacial Surgery, vol. 36, no. 2, pp. 146–152, 2007. View at: Publisher Site | Google Scholar
  6. G. R. J. Swennen, M. Y. Mommaerts, J. Abeloos et al., “The use of a wax bite wafer and a double computed tomography scan procedure to obtain a three-dimensional augmented virtual skull model,” Journal of Craniofacial Surgery, vol. 18, no. 3, pp. 533–539, 2007. View at: Publisher Site | Google Scholar
  7. T. V. Flügge, K. Nelson, R. Schmelzeisen, and M. C. Metzger, “Three-dimensional plotting and printing of an implant drilling guide: simplifying guided implant surgery,” Journal of Oral and Maxillofacial Surgery, vol. 71, no. 8, pp. 1340–1346, 2013. View at: Publisher Site | Google Scholar
  8. T. V. Flügge, S. Schlager, K. Nelson, S. Nahles, and M. C. Metzger, “Precision of intraoral digital dental impressions with iTero and extraoral digitization with the iTero and a model scanner,” American Journal of Orthodontics and Dentofacial Orthopedics, vol. 144, no. 3, pp. 471–478, 2013. View at: Publisher Site | Google Scholar
  9. C. C. Lin, C. Z. Wu, M. S. Huang, C. F. Huang, H. C. Cheng, and D. P. Wang, “Fully digital workflow for planning static guided implant surgery: a prospective accuracy study,” Journal of Clinical Medicine, vol. 9, no. 4, p. 980, 2020. View at: Publisher Site | Google Scholar
  10. H.-N. Mai, S.-Y. Choi, S.-T. Lee, and D.-H. Lee, “Optimizing accuracy in computer-guided implant surgery with a superimposition- anchor microscrew system: A clinical report,” Journal of Prosthetic Dentistry, vol. 120, no. 5, pp. 789.e1–789.e5, 2018. View at: Publisher Site | Google Scholar
  11. L. Ritter, S. D. Reiz, D. Rothamel et al., “Registration accuracy of three-dimensional surface and cone beam computed tomography data for virtual implant planning,” Clinical Oral Implants Research, vol. 23, no. 4, pp. 447–452, 2012. View at: Publisher Site | Google Scholar
  12. D. H. Lee, “Strategic use of microscrews for enhancing the accuracy of computer-guided implant surgery in fully edentulous arches: a case history report,” International Journal of Prosthodontics, vol. 31, no. 3, pp. 262-263, 2018. View at: Publisher Site | Google Scholar
  13. M. Vercruyssen, T. Fortin, G. Widmann, R. Jacobs, and M. Quirynen, “Different techniques of static/dynamic guided implant surgery: modalities and indications,” Periodontology 2000, vol. 66, no. 1, pp. 214–227, 2014. View at: Publisher Site | Google Scholar
  14. A. Almukhtar, X. Ju, B. Khambay, J. McDonald, and A. Ayoub, “Comparison of the accuracy of voxel based registration and surface based registration for 3D assessment of surgical change following orthognathic surgery,” PLoS One, vol. 9, no. 4, p. e93402, 2014. View at: Publisher Site | Google Scholar
  15. F. A. Rangel, T. J. J. Maal, M. J. J. de Koning, E. M. Bronkhorst, S. J. Bergé, and A. M. Kuijpers-Jagtman, “Integration of digital dental casts in cone beam computed tomography scans-a clinical validation study,” Clinical Oral Investigations, vol. 22, no. 3, pp. 1215–1222, 2018. View at: Publisher Site | Google Scholar
  16. H. H. Lin, W. C. Chiang, L. J. Lo, S. Sheng-Pin Hsu, C. H. Wang, and S. Y. Wan, “Artifact-resistant superimposition of digital dental models and cone-beam computed tomography images,” Journal of Oral and Maxillofacial Surgery, vol. 71, no. 11, pp. 1933–1947, 2013. View at: Publisher Site | Google Scholar
  17. T. Flügge, W. Derksen, J. Te Poel, B. Hassan, K. Nelson, and D. Wismeijer, “Registration of cone beam computed tomography data and intraoral surface scans - a prerequisite for guided implant surgery with CAD/CAM drilling guides,” Clinical Oral Implants Research, vol. 28, no. 9, pp. 1113–1118, 2017. View at: Publisher Site | Google Scholar
  18. D.-H. Lee, S.-Y. An, M.-H. Hong, K.-B. Jeon, and K.-B. Lee, “Accuracy of a direct drill-guiding system with minimal tolerance of surgical instruments used for implant surgery: a prospective clinical study,” The Journal of Advanced Prosthodontics, vol. 8, no. 3, pp. 207–213, 2016. View at: Publisher Site | Google Scholar
  19. F. Z. Jamjoom, D. G. Kim, D. J. Lee, E. A. McGlumphy, and B. Yilmaz, “Effect of length and location of edentulous area on the accuracy of prosthetic treatment plan incorporation into cone-beam computed tomography scans,” Clinical Implant Dentistry and Related Research, vol. 20, no. 3, pp. 300–307, 2018. View at: Publisher Site | Google Scholar
  20. F. Z. Jamjoom, B. Yilmaz, and W. M. Johnston, “Impact of number of registration points on the positional accuracy of a prosthetic treatment plan incorporated into a cone beam computed tomography scan by surface scan registration: an in vitro study,” Clinical Oral Implants Research, vol. 30, no. 8, pp. 826–832, 2019. View at: Publisher Site | Google Scholar
  21. S. Akyalcin, D. J. Dyer, J. D. English, and C. Sar, “Comparison of 3-dimensional dental models from different sources: diagnostic accuracy and surface registration analysis,” American Journal of Orthodontics and Dentofacial Orthopedics, vol. 144, no. 6, pp. 831–837, 2013. View at: Publisher Site | Google Scholar
  22. X. Lin, T. Chen, J. Liu, T. Jiang, D. Yu, and S. G. F. Shen, “Point-based superimposition of a digital dental model on to a three- dimensional computed tomographic skull: an accuracy study in vitro,” British Journal of Oral and Maxillofacial Surgery, vol. 53, no. 1, pp. 28–33, 2015. View at: Publisher Site | Google Scholar
  23. M. D. Scherer, “Presurgical implant-site assessment and restoratively driven digital planning,” Dental Clinics of North America, vol. 58, no. 3, pp. 561–595, 2014. View at: Publisher Site | Google Scholar
  24. C. E. Cann, “Quantitative CT for determination of bone mineral density: a review,” Radiology, vol. 166, no. 2, pp. 509–522, 1988. View at: Publisher Site | Google Scholar
  25. M. R. Norton and C. Gamble, “Bone classification: an objective scale of bone density using the computerized tomography scan,” Clinical Oral Implants Research, vol. 12, no. 1, pp. 79–84, 2001. View at: Publisher Site | Google Scholar
  26. O. Watzke and W. A. Kalender, “A pragmatic approach to metal artifact reduction in CT: merging of metal artifact reduced images,” European Radiology, vol. 14, no. 5, pp. 849–856, 2004. View at: Publisher Site | Google Scholar
  27. H.-Y. Feng, Y. Liu, and F. Xi, “Analysis of digitizing errors of a laser scanning system,” Precision Engineering, vol. 25, no. 3, pp. 185–191, 2001. View at: Publisher Site | Google Scholar
  28. A. Suomalainen, T. Vehmas, M. Kortesniemi, S. Robinson, and J. Peltola, “Accuracy of linear measurements using dental cone beam and conventional multislice computed tomography,” Dentomaxillofacial Radiolog, vol. 37, no. 1, pp. 10–17, 2008. View at: Publisher Site | Google Scholar
  29. F. Teraoka and J. Takahashi, “Dimensional changes and pressure of dental stones set in silicone rubber impressions,” Dental Materials, vol. 16, no. 2, pp. 145–149, 2000. View at: Publisher Site | Google Scholar
  30. A. Di Fiore, P. Vigolo, L. Graiff, and E. Stellini, “Digital vs conventional workflow for screw-retained single-implant crowns: a comparison of key considerations,” The International Journal of Prosthodontics, vol. 31, no. 6, pp. 577–579, 2018. View at: Publisher Site | Google Scholar

Copyright © 2020 Hai Yen Mai and Du-Hyeong Lee. This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.