Research Article

Biomechanical Comparison of a New Memory Compression Alloy Plate versus Traditional Titanium Plate for Anterior Cervical Discectomy and Fusion: A Finite Element Analysis

Table 4

The maximum von Mises stresses at the adjacent endplates in the three groups (MPa).

GroupFlexionExtensionLeft bendingRight bendingLeft rotationRight rotation

C5 inferior endplateGroup111.6414.5713.3011.479.767.41
Group216.5829.3918.6716.5011.9028.31
Group330.2353.3341.1035.1216.5552.13
C6 superior endplateGroup14.769.287.285.887.814.48
Group25.0114.089.2511.0229.4310.31
Group318.6721.7621.9119.5451.8315.57

Note: Group1 refers to the intact group, Group2 refers to the traditional plate group, and Group3 refers to the new prosthesis.